Cal Test Electronics Elditest CT2982B Скачать руководство пользователя страница 5

800-572-1028                                            caltestelectronics.com                                                            4

1  Introduction

Overview

The Elditest CT2982B is a 10 kV High Voltage divider probe for use 

with both digital and analog oscilloscopes. 

Features

 

z

Measure up to: 10 kV (DC + ACpk) or 7 kVrms AC

 

z

40 MHz bandwidth

 

z

Voltage dividing of 1000:1

 

z

Frequency compensation 

 

Содержание Elditest CT2982B

Страница 1: ...CT2982B User Manual...

Страница 2: ...ategory I CAT I only Do not exceed 60 seconds of contact with high voltage When making contact between this probe and any circuit with voltages 8 kV DC 5 kVrms AC or 15 kVpk pk AC wait for 5 minutes b...

Страница 3: ...liquid being spilled on the instrument z z In air temperatures exceeding the specified operating temperatures z z In atmospheric pressures outside the specified altitude limits or where the surroundin...

Страница 4: ...or the entire product Compliance Statements Disposal of Old Electrical Electronic Equipment Applicable in the European Union and other European countries with separate collection systems This product...

Страница 5: ...Overview The Elditest CT2982B is a 10 kV High Voltage divider probe for use with both digital and analog oscilloscopes Features z z Measure up to 10 kV DC ACpk or 7 kVrms AC z z 40 MHz bandwidth z z V...

Страница 6: ...le chassis ground 5 Before turning on the high voltage source be certain the operator is not touching any part of the device under test 6 Turn on the high voltage source 7 You will now be able to anal...

Страница 7: ...whenever the probe is connected to or transferred between oscilloscopes The procedure is the same as for a x10 passive probe z Connect the probe to the oscilloscope z Connect probe tip to square wave...

Страница 8: ...1028 caltestelectronics com 7 Figure 2 shows how over and under compensated pulse responses will look Figure 2 Over Under Compensation Figure 3 shows a perfectly compensated probe Figure 3 Good Compe...

Страница 9: ...efficient 200 ppm C Input Resistance 100 M 5 Rise Time 8 8 ns Accuracy DC Volts 3 below 10 kV AC Volts 3 at 1 kHz 3 dB 0 to 40 MHz General Cable Length 6 6 ft 2 m Designed for Use in Pollution Degree...

Страница 10: ...frequency signals be sure to comply with the Voltage vs Frequency Derating Curve CT2982B Maximum Working Voltage Derating Curve Voltage DC AC pk Frequency 100V 1KHz 10KHz 100KHz 1MHz 10MHz 100MHz 1KV...

Страница 11: ...ginal purchase Under this warranty Cal Test Electronics is limited to repairing the defective device when returned to the factory shipping charges prepaid within the warranty period Units returned to...

Страница 12: ...representative of Cal Test Electronics is authorized to assume any other obligation in connection with the sale and purchase of this device Service If you have a need for calibration or repair service...

Страница 13: ...800 572 1028 caltestelectronics com 12 Notes...

Страница 14: ...800 572 1028 caltestelectronics com 13...

Страница 15: ...800 572 1028 caltestelectronics com 14...

Страница 16: ...800 572 1028 caltestelectronics com 15...

Отзывы: