* A 1mm nylon spacer is also included with each unit. This may be
required on the camera output mount, particularly when mounting a
shutter in addition to the OptoSplit.
Having fitted the camera to the output port of the
Optosplit II it should be possible to adjust the image
to see a sharp picture of the aperture edges with the
sample in focus. The image should line up with the
edges of the aperture and should not have any
rotation or slant in either direction.
During manufacture the Optosplit II is calibrated
with the two images approximately super imposed,
so when it is first installed a single central image is
seen on screen.
3
Operation of the Optosplit II
The Optosplit II uses a single control for adjusting image seperation
and allows for different sizes of samples to be used. There are
additional controls for refining the ROI, and centering the image.
V1
V2
H2
The image should
not have any slant.
Image.
Image.
Split adjuster
V2 adjuster
H2 adjuster
V1 adjuster
Aux. component
mounts
TRIM
Clamp Screw
Trim Controller
Split adjuster
Page 6