PN MN234 C | User Manual    7 - ENG

Liberator

2. Use the following chart as a guideline to the 

recommended tubing length.

FLOW  

SETTING

MAXIMUM (RECOMMENDED)  

TUBING LENGTH*

(LPM)

20-psig

50-psig 

1-6

100 Ft. (30.5 m)

100 Ft. (30.5 m)

8

100 Ft. (30.5 m)

75 Ft. (22.9 m)

10

50 Ft. (15.2 m)

50 Ft. (15.2 m)

12

25 Ft. (7.6 m)

50 Ft. (15.2 m)

15

25 Ft. (7.6 m)

25 Ft. (7.6 m)

*Length is oxygen tubing only. Does 

not

 include a 7 Ft. cannula.

3. Verify functionality of the gauge
• Depress button to display level. If level is 

displayed and Low Battery Indicator is not il-

luminated, battery level is acceptable.

4. To verify the level of liquid oxygen in the unit 

with the liquid level gauge:

• Depress the push button on top of the unit for two 

seconds minimum. Read the LED to indicate 

contents level.

Caution: The Liberator is empty if only 
the first red LED is lit.

• If the Low Battery Indicator lights up when the 

button is depressed, inform your health care pro-

vider the next time your Liberator is filled. 

Gen 4 Meter

5. Install the DISS extension.
6. Either 
  a. Attach cannula to the DISS adapter barb on 

the DISS connection provided by health care 

provider or

  b. Attach a humidifier bottle to the DISS connec-

tion provided by health care provider:

  • Fill the humidifier bottle with distilled water 

to the proper level as indicated in the humidifier 

instructions.

  • Attach your breathing cannula to the oxygen 

tube connector on the humidifier.

7. Turn the flow control knob clockwise until the 

prescribed flow rate (numeral) is visible in the 

knob “window” and a positive detent is felt.

Caution: The knob should not be set 
higher than the maximum prescribed 
flow rate. Out-of specification oxygen 
flow will result if the flow control knob is 
set between flow rates. An indication of 
oxygen flow is the presence of bubbles 
in the humidifier bottle.
Caution: To ensure proper flow rate, 
verify fittings are tight and leak free.

Содержание Liberator 30

Страница 1: ...Liberator User Manual English...

Страница 2: ...tive 93 42 EEC concerning medical devices Authorized representative in the European Community This device complies with the requirements of Directive 93 42 EEC concerning medical devices It bears the...

Страница 3: ...kg 47 5 lb 21 54 kg 50 lb 22 68 kg 55 lb 24 95 kg 75 4 lb 34 19 kg Weight Filled 89 7 lb 40 69 kg 122 77 lb 55 68 kg 141 01 lb 63 98 kg 165 32 lb 74 99 kg 220 68 lb 100 1 kg Height 24 5 in 622 mm 29 5...

Страница 4: ...bite Liquid and gaseous oxygen though nonflammable cause other materials to burn faster than normal This hazard along with the low temperature of liquid oxygen warrants certain safety precau tions Kee...

Страница 5: ...tor is offered as a top fill model or a dual top fill and side fill unit They are filled by your health care provider The portable provides an am bulatory source of oxygen for an extended period of ti...

Страница 6: ...DOES NOT SEPARATE EASILY DO NOT USE FORCE THE UNITS MAY BE FROZEN TOGETHER LEAVE THE UNITS CONNECTED AND WAIT UNTIL THEY WARM UP THEN THEY WILL SEPARATE EASILY DO NOT TOUCH ANY FROSTED PARTS Caution S...

Страница 7: ...Indicator lights up when the button is depressed inform your health care pro vider the next time your Liberator is filled Gen 4 Meter 5 Install the DISS extension 6 Either a Attach cannula to the DIS...

Страница 8: ...utton to display level If level is displayed and Low Battery Indicator is not il luminated battery level is acceptable If the Low Battery Indicator lights up when the contents button is depressed call...

Страница 9: ...to time to relieve excess pressure especially soon after filling If hissing is persistent or abnormal this could indicate excess pressure being vented off or a leak in the system Contact service prov...

Страница 10: ...local city or town offices for instructions on proper disposal of the battery WEEE and RoHS This symbol is to remind the equipment owners to return it to a recycling facility at the end of its life pe...

Страница 11: ...tic establishments and those directly connected to the Voltage fluctuations public low voltage power supply network that supplies flicker emissions Not applicable buildings used for domestic purposes...

Страница 12: ...d device IEC 610004 4 1 kV for Not applicable input output linesNo data input output lines 1 kV line s Surge to line s Not Applicable Not Applicable IEC 61000 4 5 2 kV line s DC powered device to eart...

Страница 13: ...rya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequ...

Страница 14: ...14 ENG PN MN234 C User Manual Liberator Notes...

Страница 15: ...PN MN234 C User Manual 15 ENG Liberator Notes...

Страница 16: ...and with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA...

Отзывы: