PN MN234 G | User Manual 11 - ENG
Liberator
Safety
WARNING: PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS SUCH AS
ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER THAN 30 CM (12
INCHES) TO ANY PART OF THE LIBERATOR, INCLUDING CABLES SPECIFIED BY THE MANU-
FACTURER. OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD
RESULT.
WARNING: USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN THOSE SPECIFIED OR
PROVIDED BY THE MANUFACTURER OF THIS EQUIPMENT COULD RESULT IN INCREASED ELECTROMAG-
NETIC EMISSIONS OR DECREASED ELECTROMAGNETIC IMMUNITY OF THIS EQUIPMENT AND RESULT IN
IMPROPER OPERATION.
WARNING: USE OF THIS EQUIPMENT ADJACENT TO OR STACKED WITH OTHER EQUIPMENT SHOULD
BE AVOIDED BECAUSE IT COULD RESULT IN IMPROPER OPERATION. IF SUCH USE IS NECESSARY, THIS
EQUIPMENT AND THE OTHER EQUIPMENT SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERAT-
ING NORMALLY.
Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in this manual.
Caution: Portable and mobile radio frequency (RF) communications equipment can affect
Medical Electrical Equipment.
Caution: The Liberator Reservoir should not be used adjacent to or stacked with other equip-
ment and that if adjacent or stacked use is necessary, the Liberator Reservoir should be
observed to verify normal operation in the configuration in which it will be used.
Table 1
Guidance and Manufacturer’s declaration—electromagnetic emissions
The Liberator is intended for use in the electromagnetic environment specified below. The customer or
the user of the Liberator should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic environment—guidance
RF emissions
Group 1
The Liberator uses RF energy only for internal function.
CISPR 11
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B
Harmonic emissions The Liberator is suitable for use in all establishments, includ-
ing
IEC 61000-3-2
Not applicable domestic establishments and those directly connected to the
Voltage fluctuations/ public low-voltage power supply network that supplies
flicker emissions
Not applicable buildings used for domestic purposes.
IEC 61000-3-3
Содержание Liberator 20