FreeStyle Comfort
0x0200
Breath Rate
Exceeded
Patient
Breathing more
than 40 BPM
Other
Run unit in normal operation
for 15 minutes without
breathing off the unit.
Ensure alarm does not occur
during this period.
Contact Caire to send the
unit in for repair.
0x0400
Fan Failure
Other
Contact Caire to send the
unit in for repair.
0x0800
Low
Pressure
Intake Filter
Clogged
Pinched or
Clogged Inlet
Tubing
Feed/Waste
Valve Failure
Sieve Bed
Failure
Other
Clean or replace intake
filters.
Verify that compressor inlet
tubing is not pinched or
clogged.
Replace feed/waste valves.
Check for leaks on bottom
sieve bed caps. Replace
sieve beds.
Contact Caire to send the
unit in for repair.
0x1000
Low
Temperature
Case
Temperature
less than 0 °C
Allow unit to warm up to
room temperature
0x2000
Compressor
Failure
Other
Contact Caire to send the
unit in for repair.
0x4000
O
2
Out of
Range
Other
Contact Caire to send the
unit in for repair.