SERVICE MODE FUNCTIONS

 

Factory maintenance or service updates may sometimes be required on the Eclipse 5. Factory and qualified 
factory-trained technicians can access service mode software functions by using the Service Port located on 
the back of the unit. The Service Port is not for patient use.

 

 

 

Figure 4: Eclipse 5 Service Port

 

 
 
 
 
 
 
 
 

 

ECLIPSE 5 DATA ACQUISITION TOOL (EDAT)

 

EDAT 

is the world’s first computer-based data retrieval service tool for oxygen concentrators. EDAT is proprietary 

to CAIRE Inc., and is comprised of both hardware and software components. EDAT is CAIRE 

Inc.’s global solution 

for your service and support needs.

 

 

EDAT software is capable of sensing and displaying historical event logs and values of components and sensors 
within the Eclipse 5 system. This can be used to determine system faults or user errors and communicate data 
for troubleshooting or documentation related to service and updates. EDAT hardware attaches or Plugs into the 

Eclipse 5 Service Port, and into a computer’s USB port.

 

 

The EDAT software set allows a provider to set-up a hub and spoke service center. Field service reps can travel 

to a patient’s home, troubleshoot and transmit the data to either in-house tech support personnel or CAIRE Support 
personnel can diagnose, provide solutions and print documentation for equipment records.

 

 

EDAT also makes Eclipse 5 software upgrades easy and accessible.

  

 

For more information on EDAT, contact CAIRE Inc. Technical Service at 1-800-482-2473 and reference EDAT PN 5535-SEQ.

 

 

Содержание Eclipse 5 with autoSAT

Страница 1: ...Eclipse 5 with autoSAT Oxygen Concentrator TECHNICAL SERVICE MANUAL...

Страница 2: ...AC Power Supply 18 DC Power Supply 18 Power Cartridge 19 Typical New Power Cartridge Operation Time 20 Charging Algorithm 22 Training The Patient 23 Introduction 23 Pre Delivery Check List 23 Indicati...

Страница 3: ...ication Tests 48 Record Hours of Operation Software Version 49 Electrical Safety Test 50 Cleaning the Eclipse 5 50 Provider Service and Maintenance Record 51 Shipping and Transporting the Eclipse 5 52...

Страница 4: ...rinciples of operation Providers are given proper guidance in the use of the Eclipse 5 and its accessories that can beconveyed to patients Providers are made aware of the care diagnostics maintenance...

Страница 5: ...rd commercial aircraft Flow Setting Indicator Pulse Mode Operation Device operating normally power button Increase Flow Setting Decrease Flow Setting ISO 7000 Graphical symbols for use on equipment In...

Страница 6: ...RIDGE BATTERY PN 7082 SEQ PN 20852326S with IEC type A cord 20852325S with IEC type C cord and 20852324S with IEC type G cord two LED model Oxygen Outlet Port Control Panel FAA Approval Icon Rating La...

Страница 7: ...0 LPM Continuous Flow 3 0 Pulse Dose Mode 48 dB A 40 dB A Less than 57 dB while alarming Operating Environment Temperature Humidity 50 F to 104 F 10 to 40 C 10 to 95 Non condensing 82 4 F 28 C Maximu...

Страница 8: ...w at the last Continuous Flow setting After another 15 seconds the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a breath The Eclipse 5 will stay in this modality until a breath...

Страница 9: ...gnetic Compatibility Emissions EN55011 FCC Part J15 Class B Conducted EN55011 FCC Part J15 Class B Radiated EN61000 3 2 EN61000 3 3 EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 5 EN61000 4 6 EN61000...

Страница 10: ...provided in this manual Portable and mobile RF communications equipment can affect Medical Electrical Equipment The use of Accessories transducers and cables other than those specified with the excep...

Страница 11: ...ons it is recommended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnetic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels char...

Страница 12: ...nd controls both the power source and the Power Cartridge charger Figure 1 Eclipse 5 System Block Diagram ATF CONCENTRATOR MODULE The Eclipse 5 uses a passive system to separate oxygen from air Air fl...

Страница 13: ...sor enclosure serves to efficiently draw cooling air in over the compressor cylinders while simultaneously pushing exhaust gas out of the concentrator POWER DISTRIBUTION The Power Manager takes extern...

Страница 14: ...ng has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approxi...

Страница 15: ...y rates Regardless of setting the pulse profile is a simple square wave pulse based on a 16 mL volume Refer to Pulse Profiles table below Pulse durations are no less than 100 milli seconds at the begi...

Страница 16: ...S x PS 1 is the most sensitive setting PS 3 is the least sensitive setting Factory default is PS 2 HOURS OF OPERATION The Eclipse 5 hour meter provides valuable information on hours of operation Provi...

Страница 17: ...t needs EDAT software is capable of sensing and displaying historical event logs and values of components and sensors within the Eclipse 5 system This can be used to determine system faults or user er...

Страница 18: ...t is illuminated when the AC Power Supply is supplying 28VDC power If the LED is not illuminated there is no input power available In addition the Power Supply contains protection circuits for output...

Страница 19: ...f the user If the Eclipse 5 power cartridge is being charged the battery symbol will display with a waterfall animation However if the power cartridge is not charging then the battery symbol will rema...

Страница 20: ...d with the Eclipse 5 is not fully charged Before using the Eclipse 5 Oxygen System for the first time the Power Cartridge must be fully charged Store the power cartridge in a cool dry location Do not...

Отзывы: