background image

 

 

 

 

 

 

 

 

 

 

Document type: 

Title : 

Re vision date: 

Re vision: 

User's Manual (MUT) 

Mod. A1735-A1835-A1735B-A1835B HV Boards 

24/03/2014 

 
 
 

NPO : 

File name: 

Number of pages: 

Page : 

00101/00:17XXx.MUTx/08 

A1735B-A1835B_REV8.DOCX 

14 

 

1.4 

Technical Specifications 

 

1.4.1 

Packaging 

 
The Mod. A 1735 –  A 1835 modules are housed in a 5 TE-wide, 6U-high mechanics. The 
Mod. A 1735B – A 1835B  modules are housed  in a 10 TE-wide, 6U-high mechanics. 

1.4.2 

External connections 

The location of all components of the front panel is shown in Fig. 2.1, p. 8. The function 
and electro-mechanical specifications of the external connectors are listed in the 
following subsections. 
 

CH  0..11 (A 1735  – A 1835): 

Mechanical specifications:

 

HV coaxial connectors  

  Radiall SHVR317580-type 

Electrical specifications

: high  voltage  output 

according  to specifications  given  in Table 
2.1, p.6. 

CH  0..27 (A 1735B  – A 1835B): 

Mechanical specifications:

 

Multipin connector  Radiall 691803004-type, 
52  pin male (to be mated with Radiall 
691802002-type

5

  [

SCEM 09.41.34.700.2

]); 

see Fig. 2.2 for  pin assignment.    

Electrical specifications

: high  voltage  output 

according  to specifications  given  in Table 
2.1, p.6. 

HV EN CONNECTOR 

Mechanical specifications

 

 

00-type LEMO connector. 

Electrical specifications:

 board ENABLE 

input, if connected to ground,  the channels 
are enabled. Refer to § 3.2, p.13  for further 
details. 
 

                                                                 

5

  Requires 52 pins Radiall 691804300 [

SCEM 09.41.33.830.7

]

 

type, to be installed using the insertion/extraction tool 

Radiall  282549024 [

SCEM 34.95.17.125.3

] type. 

Содержание A 1735

Страница 1: ...Technical Information Manual MOD A 1735 A 1835 A 1735B A 1835B 12 28 CHANNEL HV BOARDS MANUAL REV 8 24 March 2014 Revision n 8 NPO 00101 00 17XXx MUTx 08...

Страница 2: ...ight to change partially or entirely the contents of this Manual at any time and without giving any notice Disposal of the Product The product must never be dumped in the Municipal Waste Please check...

Страница 3: ...ON 5 1 2 CHANNEL CHARACTERISTICS TABLE 6 1 3 FRONT PANEL 7 1 4 TECHNICAL SPECIFICATIONS 8 1 4 1 Packaging 8 1 4 2 External connections 8 1 4 3 Displays 9 1 4 4 Other components 9 2 SAFETY INFORMATION...

Страница 4: ...5B A1835B_REV8 DOCX 14 4 LIST OF FIGURES FIG 1 1 MOD A 1735 A 1835 LEFT AND MOD A 1735B A 1835B RIGHT FRONT PANEL 7 FIG 1 2 MOD A 1735B A 1835B HVMULTIPIN CONNECTOR 9 FIG 3 1 A MOD A1735 A 1835 SIDE V...

Страница 5: ...DOWN rates may be selected independently for each channel in the range 1 500 V s in 1 V s steps The output current is monitored with 20 nA 500 nA resolution depending on current range if a channel tri...

Страница 6: ...nA depending on current range Mod A 1835 A 1835B VMAX hardware 0 1 5 kV common for all the board channels VMAX hardware accuracy 2 of FSR VMAX software 0 1 5 kV settable for each channel VMAX softwar...

Страница 7: ...T P U T S SCALER POSITIVE 12 CH Mod V560E Mod A 1835P HV EN input connector Channel output connector Channel LED SCALER Mod V560E Mod A1835BP 28 CH HV EN MAXV GND GND GND GND GND NC NC NC NC NC 6 NC...

Страница 8: ...35 A 1835 Mechanical specifications HV coaxial connectors Radiall SHVR317580 type Electrical specifications high voltage output according to specifications given in Table 2 1 p 6 CH 0 27 A 1735B A 183...

Страница 9: ...positive polarity version yellow LEDs for negative polarity version HV ON LED A 1735B A1835B Function it lights up if at least one output channel is on Type red LED for positive polarity version yell...

Страница 10: ...ck or fire hazard do not apply a voltage to a load that is outside the range specified for that load Avoid Electric Shock To avoid injury or loss of life do not connect or disconnect cables while they...

Страница 11: ...product The following symbols may appear on the product DANGER ATTENTION High Voltage Refer to Manual 2 3 Installation The Mod A 1735 A 1835 are single width boards which can be inserted in any slot...

Страница 12: ...tput channel it is possible through the SYX527 system to perform the following operations Assign to channel a symbolic name Set output voltage VSET Set max output current ISET Set output voltage softw...

Страница 13: ...o enable the board s operation Provide the HV EN connector with a 50 Ohm termination Provide short circuit between Pin 51 and Pin 52 IINT and IEXT of the HV Multipin Connector see Fig 2 2 this is a sa...

Страница 14: ...els Dip Switch side view T A B C 4 3 2 1 A B Dip Switch T A B C 4 3 2 1 Channels Channels Channels Channels Channels Channels Channels Fig 3 1 A Mod A1735 A 1835 side view B Mod A1735B A 1835B top vie...

Отзывы: