easy2log© RT0005 Technical Information Manual
11
Manual interface behaviour
The easy2log
©
RT0005 tag has a simple manual interface that includes a button, a green LED and a red LED that,
depending on the configuration, permits to check the status of the alarms and to start/stop the logging activity. By
default, the button can be used only to check the status of the tag but not to start/stop the logging activity. Start/stop
function can be activated setting the Button Enable (BE) bit in the CONTROL register.
The manual interface behaves as follows:
Pressing the button quickly (less than 2 second) activates the check status function:
If the battery level is still sufficient for the logging activity and no alarms are active the green LED will stay lit for 2
seconds.
If the battery level is still enough for the logging activity but there are active alarms the red LED will stay lit for 2
seconds. The alarms causing the red LED to ligth are:
Memory full (see MEMFULL bit in STATUS register description).
ETA alarm (see ETA_ALRM bit in the STATUS register description).
Bin counters alarm (see BIN_ALMR bit in STATUS register description).
Mean Kinetic alarm (see MKT_ALRM bit in the STATUS register description).
Shelf life alarm (see SHL_ALRM bit in the STATUS register description).
If the battery has not enough charge to continue the logging activity the red or green LED will blink for 2 seconds (the
green one in the case of no alarms, red in case of some alarms).
NO ALARM
ALARM
Temperature Status
Green
Red
NO ALARM
ALARM
Battery Status
No flashing
Flashing
Tab. 2.1: Alarm indicator
If the BE bit is 1 (Start/stop function activated), in addition to the behaviour previously described, the manual interface
has the following features:
The start of the logging activity can be performed in the following way:
Hold down the button
Wait until the green LED lit and the red LED start blinking
Release and press again the button within 5 seconds to confirm the command otherwise the
command is ignored.
If the command is accepted the green LED will be lit for 15 seconds.
The acquisition is started.
And, if the SD bit is 0, also the following operation is allowed:
The stop of the logging activity can be performed in the following way:
Hold down the button
Wait until the red LED lit and the green LED start blinking
Release and press again the button within 5 seconds to confirm the command otherwise the
command is ignored.
If the command is accepted the red LED will be lit for 15 seconds.
The acquisition is stopped.
Содержание Easy2log RT0005
Страница 1: ...RT0005 UHF Semi Passive Logger Tag Technical Information Manual Revision n 02 12 03 2013...
Страница 33: ...easy2log RT0005 Technical Information Manual 33 Mechanical Specification...
Страница 34: ...easy2log RT0005 Technical Information Manual 34 5 RT0005 Temperature Measurement Traceability...