![Caen ELS FAST-Bi-1K5 Series Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/caen-els/fast-bi-1k5-series/fast-bi-1k5-series_user-manual_4315014031.webp)
FAST-Bi-1K5 – User’s Manual
31
1.
Ensure that Mains switch is on Off position “O”
2.
Remove factory jumpers between
+S
to
+
and
–S
to
–.
3.
Using a twisted pair or shielded cable (suggested wire size is 0.3 or 0.5 mm
2
)
connect the
+S
terminal to the positive output terminal and the
–S
to the
negative output terminal as illustrated in
4.
For FAST-Bi-1K5 with output voltage rating > 50 V,
fix the Safety cover to
the rear panel using the M3x10 mm screw,
Figure 16
: Remote Sensing Cover
In order to perform remote sensing at different points – e.g. the load terminals
– it would be necessary to connect Pin #1 and Pin #4 as in
Figure 17
: Example of Remote Sensing
3.3
Persistent Switch
The FAST-Bi-1K5 is equipped with a Persistent Current Switch (PCS) output
) capable of sourcing up to 240 mA. The PS is shipped with
the terminal block plug from Wurth-Elektronik (P/N: 691304330002). The PCS output
current can be adjusted by variable resistor indicated as “I ADJ” in
. The PS
is shipped with a PCS output current configured at 50 mA.
The PCS output can be controlled from the GUI or with specific PS commands
(refer to the
“Command Reference Manual”
for more information).
FAST-Bi-1K5
L
OAD
+S
-S
+
-
Power Cables
Sensing Cables
OUTPUT