The Flow % is a value that need to be set at the start-up in the CC50 or CC200 controller. 
To have your exact Max. Value, you need to multiply the Max Value when the Flow % = 
100, with the real Flow % of the operational system.

 

 

EX :   Flow % = 3200 CFM/ 4000 CFM = 80 %. 

  

Max Value Bag filter = 144 x 80% = 115 = 1.15'' h2o. 

 

 

REPLACEMENT FILTER EQUIVALENTS  

 

PREFILTERS:  MERV7 (30% ASHRAE 52-76) - ULC Class II  

American Air Filter:   

 

 

24” x 24” x 2” - AM-AIR Class II    

 

 

12” x 24” x 2”  - AM-AIR Class II  

 

BAG FILTERS:  MERV15 (90 - 95% ASHRAE 52 

– 76) - ULC Class II  

American Air Filter:       

 

 

24” x 24” x 21” - DRI-PAK - Class II                  

 

 

12” x 24” x 21” - DRI-PAK - Class II  

 

BOX FILTERS:  MERV16 (95% DOP/99% ASHRAE 52-76) ULC Class II  

American Air Filter:                    

 

 

24” x 24” x 12” - BIOCELL Class II                    

 

 

12” x 24” x 12” - BIOCELL Class II Farr Filter:                    

 

ODOR  MEDIA:    1/8”  Activated  alumina  pellets  impregnated  with  potassium 
permanganate.  

American Air Filter:      

Permasorb  

Содержание Stadium Series

Страница 1: ...Stadium hood Series Maintenance operation Manual Version V1 0...

Страница 2: ...ypically most micron and submicron particles escape into the exhaust ductwork The exhaust ducting is connected from the hood to the inlet of the Stadium Serie This exhaust ductwork must be supplied an...

Страница 3: ...ent of a high temperature in the ductwork leading to the Stadium unit a firestat located at the inlet of the Stadium unit filter section is activated When the exhaust air reaches 160 F the firestat is...

Страница 4: ...r resistance through each filter increases Even though the Stadium unit has a heavy duty Class II backward inclined fan the increase in combined resistance static pressure W C through each filter will...

Страница 5: ...g the unit thereby making commissioning a more straightforward process Additionally MAXFLOW eliminated the nee to change sheaves If the exhaust volume has to be field adjusted because of an appliance...

Страница 6: ...as well 2 Run the unit until the Prefilter Clogged lights turns on When the light comes on the unit will shut down Immediately turn the Override switch clockwise and put the unit into override The uni...

Страница 7: ...Filter Change No 1 Actual 2 Determine the number of days between the Startup date and the Box filter Change No Actual date Subtract two days from this number Add the this number of days to the last ac...

Страница 8: ...NTS PREFILTERS MERV7 30 ASHRAE 52 76 ULC Class II American Air Filter 24 x 24 x 2 AM AIR Class II 12 x 24 x 2 AM AIR Class II BAG FILTERS MERV15 90 95 ASHRAE 52 76 ULC Class II American Air Filter 24...

Страница 9: ......

Страница 10: ...m to alert that one of the filter is almost clogged Every Month 1 Complete the two week list 2 Inspect the exhaust fan belt for correct tension and wear All belts usually require adjustment at this ti...

Страница 11: ...ending a sample to an accredited test laboratory Most major filter suppliers have access to such service Replace media if required To replace the media remove the cells from the STADIUM unit Open the...

Страница 12: ...A noisy fan is a typical sign of a fan out of balance V Belt Drives 1 ALWAYS KEEP SPARE SET OF BELTS Periodically check the belt tension and adjust if necessary 2 Some slack should be left in the bel...