background image

1

PRODUCT 

MANUAL

OPERATING INSTRUCTIONS

CABPAT

Содержание CABPAT

Страница 1: ...1 PRODUCT MANUAL OPERATING INSTRUCTIONS CABPAT ...

Страница 2: ...lass II Appliance 11 3 4 Testing a Class II Appliance with Leakage 12 3 5 Testing a mains cord 14 3 6 Testing an extension lead 15 3 7 a Test an EPOD Powerboard 16 3 7 b Test an EPOD Extension Lead with MOV and or RCDs 17 4 Troubleshooting 20 5 Specification 21 6 Maintenance 22 7 Battery Check 23 7 1 Battery Replacement 23 8 Fuse 24 8 1 Fuse Replacement 24 9 Service and Calibration 25 ...

Страница 3: ...er to avoid danger Caution risk of electric shock Always ensure that the equipment under test is discon nected from any supply before connecting to the tester It is the responsibility of the user to ensure that the operation of the unit is correct and that the accuracy of the measurements are within specification Before use ensure unit is clean and dry visually inspect all leads con nectors and ca...

Страница 4: ...4 PRODUCT MANUAL Figure 1 CABPAT Front View Figure 2 CABPAT End View ...

Страница 5: ...nection 11 Earth Insulation Test Terminal User Interface The LCD display shows test progress results for individual tests and the overall test result for an appliance or mains cord Measurements displayed next to the RPE icon indicate the resistance of the protective earth conductor Measurements displayed next to the RISO icon indicate the resistance measured by the insulation test Measurements dis...

Страница 6: ...wer ON OFF press and hold the CL I and CL II test buttons simultaneously CABPAT will power on off when buttons are released Note The unit will automatically switch OFF after approximately 3 minutes if no keys are pressed ...

Страница 7: ...es a visual inspection proceed with the electrical tests Plug the black test lead into the Earth Insulation test terminal on the CAB PAT end panel Plug the appliance into the CABPAT front panel test socket Connect the black test probe to an exposed metal part on the appliance If the Appliance under test has an ON OFF switch make sure it is in the ON position Press the CL I test button to start a C...

Страница 8: ... and the unit indicates a pass result as shown below If the Earth Continuity test has passed then the unit will proceed with an Insulation test Note The power switch on the appliance under test must be in the ON position to perform an insulation test If no appliance is detected the CABPAT will display LO LOAD If the LO LOAd enunciator is displayed the load presented by the appli ance may be too sm...

Страница 9: ...ket When plugging the mains lead into the CABPAT ensure that the polarity orientation of the connector is correct DO NOT FORCE THE PLUG INTO THE CONNECTOR doing so may damage the CABPAT Plug the black test lead into the Earth Insulation test terminal on the CABPAT end panel Plug the appliance into the CABPAT front panel test socket Connect the black test probe to an exposed metal part on the appli...

Страница 10: ...s checked and the status is indicated using the LN LE and NE enunciators of the LCD LN LE NE Mains Status ON ON OFF Correct Flash Flash OFF No mains OFF Flash Flash Earth fault OFF Flash Flash Live neutral reserved If the LO LOAd enunciator is displayed the load presented by the appliance may be too small for the CABPAT to detect In this case press the Leakage button to continue Note If the High L...

Страница 11: ...ug the appliance into the CABPAT front panel test socket Connect the black test probe to an exposed metal part on the appliance If the Appliance under test has an ON OFF switch make sure it is in the ON position Press the CL II test button to start a Class II test sequence Note The power switch on the appliance under test must be in the ON position to perform an insulation test If no appliance is ...

Страница 12: ... Visually inspect the appliance as per requirements of AS NZS3760 If the ap pliance passes a visual inspection proceed with the electrical tests Plug the mains lead into the CABPAT and a mains supply socket When plugging the mains lead into the CABPAT ensure that the polarity orientation of the connector is correct DO NOT FORCE THE PLUG INTO THE CONNECTOR doing so may damage the CABPAT Plug the bl...

Страница 13: ...ll flash press the Leakage button to start a Leakage test The mains supply status is checked and the status is indicated using the LN LE and NE enunciators of the LCD If the LO LOAd enunciator is displayed the load presented by the appli ance may be too small for the CABPAT to detect In this case press the Leakage button to continue Note If the High Load enunciator is displayed the load presented ...

Страница 14: ...uity of the protective earth If the measured value is greater than 1 ohm a cross is placed next to the RPE enunciator a FAIL is indicated If the measured value is less than or equal to 1 ohm a tick is placed next the RPE enunciator The unit will proceed with the Insulation test If the Insulation Resistance is less than 1 Mohm a cross is placed next to the RISO enunciator If the Insulation Resistan...

Страница 15: ...s that the live and neutral connections are crossed live and neutral conductors reversed 3 6 Testing an extension lead Visually inspect the appliance as per requirements of AS NZS3760 If the cord passes a visual inspection proceed with the electrical tests Plug the supplied 0 5m IEC lead into the IEC socket and into a mains outlet on the extension lead Plug the mains plug of the extension lead int...

Страница 16: ...ry last socket Only an earth resistance test is carried out on the remaining sockets Procedure A 1 Plug powerboard lead into mains outlet socket on the CABPAT 2 Plug IEC Lead into the IEC socket on the top of the CABPAT 3 Plug the IEC Lead plug into the first socket of the EPOD 4 Press CL I test button to perform a complete test sequence covering earth insulation and polarity to test the socket 5 ...

Страница 17: ... AS NZS3760 requires an earth resistance test to be performed to each outlet This test sequence uses a leakage test to replace an insulation test at 500V DC to avoid triggering the MOV surge protector or to apply power to the RCD to allow circuit connections Operators have the choice of carrying out two test sequences for EPODs with MOV and or RCD Procedure A a complete test sequence is carried ou...

Страница 18: ...socket 7 Repeat sequence on every socket Procedure B 1 Plug the mains lead into the CABPAT and a mains supply socket 2 Plug powerboard lead into mains outlet socket on the CABPAT 3 Plug IEC Lead into the IEC socket on the top of the CABPAT 4 Plug the IEC Lead plug into the first socket of the EPOD Note First socket is defined as the closest socket to the power lead 5 Press the CL I and LKGE button...

Страница 19: ...utton to carry out a leakage and polarity test on the last socket This completes the sequence for the powerboard test Note Construction site EPODs usually have 2 banks of outlet sockets A complete test sequence must be carried out on at least one socket outlet on each bank ...

Страница 20: ...rrectly and is switched on Press the CL I button to proceed with a Class I sequence the CL II button to proceed with a Class II se quence or the Leakage button to proceed with a Leakage test Q The CABPAT shows a HIGH LOAD warning A The appliance under test is greater than 10A proceeding with the Leakage test may blow the fuse in the CABPAT It is suggested that the sequence is aborted and a sequenc...

Страница 21: ...0 19 99 Mohms Accuracy 5 2 digits Test voltage 500V VDC Test current 1mA into 500k Test current 2mA into 2k Class I Leakage Current Display Range 0 15 9 99 mA Measuring Range 0 25 9 99 mA Accuracy 5 2 digits Test voltage Mains Supply voltage Class II Leakage Current Display Range 0 10 5 00 mA Measuring Range 0 10 5 00 mA Accuracy 5 2 digits Test voltage Mains Supply voltage Cord Test Earth continu...

Страница 22: ...ge category 300V CAT II 6 Maintenance Clean only with a dry cloth do not use solvents Before use ensure unit is clean and dry visually inspect all leads connectors and case Any damage or wear must be rectified to preserve user safety Check the battery contacts and compartment are free of electrolytic contamination Any contamination of the battery contacts or compartment should be cleaned with a dr...

Страница 23: ...attery Replacement Before opening the CABPAT ensure that all test leads are disconnected Switch off the unit by pressing and holding CL I and CL II buttons Disconnect all leads from the CABPAT Place the CABPAT face down and release the captive screw in the battery compartment cover Remove the battery compartment cover and remove the discharged batteries Replacement batteries must be either Alkalin...

Страница 24: ...it by pressing and holding CL I and CL II buttons Disconnect all leads from the CABPAT Place the CABPAT face down and release the captive screw in the battery compartment cover Remove the battery compartment cover and remove the fuse Ensure that the replacement fuse is a 20mm 10A 250V F Ceramic Fitting the incorrect fuse type may lead hazard Insert the replacement fuse into the fuse holder Relocat...

Страница 25: ... advice on Service and Calibration contact AUSTRALIA Service Department Emona Instruments 78 Parramatta Rd Camperdown NSW 2050 PO Box 15 Camperdown NSW 2050 Tel 02 9519 3933 Fax 02 9550 1378 Email service emona com au Sales Offices VIC 1206 Toorak Rd Camberwell VIC 3124 Tel 03 9889 0427 QLD Unit 1 1644 Ipswich Rd Rocklea QLD 4106 Tel 07 3275 2183 SA TAS 3 26 The Parade West Kent Town SA 5067 Tel 0...

Страница 26: ...r doing a leakage test Hi Load Appliance may draw more then 10 amps Appliance under test is greater than 10A con sider doing Insulation test LOAD MESSAGES ERROR CODE MEANING ACTION err 10 Internal relay failure Return for Service err 12 No Insulation Volts Return for Service err 61 Internal relay failure Return for Service err 63 Internal relay failure Return for Service err 80 Internal relay fail...

Страница 27: ... LEAKAGE TEST POLARITY TEST CLI Class I Earthed Appliance 200mA 500V CLII Class II Double Insulated Appalance 500V CLI Lead Power Board LEAD IEC or Extension 200mA 500V Yes CLI Leakage Class I Power up Test 200mA 240VAC CLII Leakage Class II Power up Test 240VAC TEST SEQUENCE ...

Страница 28: ...lack connect ed to exposed mental parts Most hand held tools 2 or 3 pin socket Leads or Power Board Class I IEC Leads Adaptor red For power tools every socket must be checked Class I Leakage Appliance lead black connected to exposed mental parts Beware moving parts Appliance is active Class II Leakage Appliance leads black connect ed to exposed mental parts Beware moving parts Appliance is active ...

Отзывы: