- 6 -
times, and then relief it.
②
.
Press “Func” key until the indicating lamp “Verifing” flashed, press “Enter” key to start
verification. “d0000” (set load for verification) will be displayed on the Force Value window and
“XXXXX” (actual force value) will be displayed on Loading Rate window.
③
.
Verify the zero set load. Make the piston rise slowly, press “Clear” key to remove the tare when
the proving ring is close to the upper platen. Press “Enter” key for the next verification.
④
.
The second set load will be displayed on the Force Value window and the actual force value will
be shown on the Loading Rate window. Press “Enter” key when the value in the proving ring is
the same with the standard value.
⑤
.
Repeat the procedures above for the next set loads verification.
⑥
.
Print the verification report after the verification is completed.
Sample report:
standard verified
01) 0 0.00
02) 200 199.80
03) 400 399.51
04) 800 700.02
05) 1200 1199.3
06) 2000 1998.9
06-03-24 13:25
Checked by: 01
Verified by:02
If the data verified is in conformity with the requirements, the pressure controller is passed,
otherwise, it is necessary to calibrate the controller.
Calibration
It is necessary to calibrate the testing machine if the verification is not passed after the machine is
used for a period of time.