![C-LOGIC 8900 Скачать руководство пользователя страница 17](http://html2.mh-extra.com/html/c-logic/8900/8900_user-manual_3100367017.webp)
Three-phase active power (W)
Escala
Precisión
Resolución
3kVA
±(3%+5)
0.001kVA
12kVA
±(3%+5)
0.01kVA
30kVA
±(3%+5)
0.01kVA
120kVA
±(3%+5)
0.1kVA
150kVA
±(3%+5)
0.1kVA
600kVA
±(3%+5)
0.1kVA
Minimum test current: 2A; Minimum test voltage: 50V
Power factor
Escala
Precisión
Resolución
0.3~1
Capacitivo
±(0.02%+2)
0.001
0.3~1
Inductivo
±(0.02%+2)
0.001
Minimum test current: 2A; Minimum test voltage: 50V
Reactive power
Escala
Precisión
Resolución
3kVAr
±(3%+5)
0.001kVAr
12kVAr
±(3%+5)
0.01kVAr
30kVAr
±(3%+5)
0.01kVAr
120kVAr
±(3%+5)
0.1kVAr
150kVAr
±(3%+5)
0.1kVAr
600kVAr
±(3%+5)
0.1kVAr
Minimum test current: 2A; Minimum test voltage: 50V. The
reactive power is calculated according to the measured
values of V, A and kW.
AC RMS voltage
Escala
Precisión
Resolución
Impedancia de
entrada
80V
±(1%+5)
0.1V
1 MΩ // 10 pF
180V
±(1%+5)
0.01V
400V
±(1%+5)
1V
600V
±(1%+5)
1V
Maximum allowed overload voltage: 750V (RMS)
AC RMS current
Escala
Precisión
Resolución
20A
±(2%+5)
0.01A
40A
±(2%+5)
0.01A
100A
±(2%+5)
0.1A
200A
±(2%+5)
0.1A
450A
±(2%+5)
1A
1000A
±(2%+5)
1A
Maximum allowed overload current: 1200A
Single phase active power (W)
Escala
Precisión
Resolución
30kW
±(3%+5)
0.01kW
60kW
±(3%+5)
0.01kW
120kW
±(3%+5)
0.1kW
150kW
±(3%+5)
0.1kW
300kW
±(3%+5)
0.1kW
600kW
±(3%+5)
0.1kW
Minimum test current: 2A; Minimum test voltage: 50V
29
30