![BYK LC 4 Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/byk/lc-4/lc-4_operating-manual_2837038050.webp)
Technical data
BYK LC 4
50
Measuring ranges
and
resolution
Accuracy
(± 1 digit)
[µS/cm]
[mS/cm]
0.0 ... 199.9
0 ... 1999
0.00 ... 19.99
0.0 ... 199.9
Spec. resistance
[k
Ω
*cm]
Spec. resistance
[M
Ω
*cm]
0.000 ... 1.999
0.00 ... 19.99
0.0 ... 199.9
0 ... 1999
0.00 ... 19.99
0.0 ... 199.9
0 ... 1999
SAL
0.0 ... 70.0 according to the IOT table
TDS [mg/l]
0 ... 1999
Factor can be set between 0.40 ... 1.00
T [°C]
T [°F]
−
5.0 ... + 105.0
23.0 ... + 212.0
,
Specific resistance
No compensation:
Accuracy ± 0.5 %
Nonlinear compensation
:
Accuracy
Temperature of test sample
± 0.5 %
0 °C ... 35 °C
acc. to ISO 7888, EN 27888
(*)
± 0,5 %
35 °C ... 50 °C
enhanced nLF function
Linear compensation
Accuracy
Temperature of test sample
± 0.5 %
10 °C ... 75 °C
(The accuracy percentage always refers to the
measured value.)
(*)
Corresponds to DEV C8