Rig Rat III Controller
User Manual
Storage and Transporting
The controller, when used for continuous monitoring on
temporary job sites or as a roving monitor, is easily installed
and stored. Battery powered instruments can be used
intermittently or stored for long periods of time without
recharging as BW batteries have a very low rate of self-
discharge.
It is important to be aware of several factors in the storage
and transportation of a battery powered equipment.
Prior to Storage or Transporting
Follow these steps before transporting an instrument or
placing an instrument in storage:
1.
Ensure all instruments are fully charged.
2.
Turn off the instrument.
3.
Unplug all equipment from the controller and
replace port cover caps. Keep a record of all
connections.
4.
Loosely coil and secure all cables to prevent
damage.
Note
Controllers and rechargeable options should be
recharged at least once every six months when in
storage.
a
Caution
Completely discharged batteries should not be
left longer than four weeks before recharging. If
the battery is fully discharged for a long period
of time, cycle charge the unit several times to
restore function. Dependent on the conditions,
cycle charging may restore up to 75% of a full
charge. Ensure all equipment is fully charged
before returning to service.