GasAlertMicro 5/PID/IR
User Manual
7KHGHWHFWRUVDYHVWKHFDOLEUDWLRQDQGUHWXUQVWRQRUPDORSHUDWLRQ
Auto Span
:KHQDXWR]HURLVFRPSOHWHWKHIROORZLQJVFUHHQVGLVSOD\
.Note
Span sensors in the following order:
•
Exotics (NH
3
, ClO
2
, O
3
, Cl
2
, and CO
2
)
•
Single gas
•
Quad gas (H
2
S, CO, LEL, and O
2
)
•
PID
Note
A generator must be used for O
3
and ClO
2
sensors.
To ensure accurate calibration, a single gas calibration cap
must be used to calibrate O
3
and ClO
2
sensors.
•
&O2
Use a Tedlar bag as a buffer between the generator
and the detector (using the single gas calibration cap) to
regulate the flow rate to ensure accurate readings.
Allow the Tedlar bag to fill for several minutes before
initiating calibration.
•
2
Calibrate using the generator and the single gas
calibration cap. Do not use a Tedlar bag.
Set the generator to 0.5 ppm at a flow rate of 0.5 l/min.
'HSHQGLQJXSRQWKHJDVF\OLQGHUEHLQJXVHGRQHRUDOOIRXUVHQVRUV
FDQEHFDOLEUDWHGDWRQHWLPH
$SSO\6SDQ*DV1RZ
*DV$OHUW0LFURDQG3,'
*DV$OHUW0LFUR,5
www.
CanarySense
.ca
Shop for Gas products online at:
1.800.561.8187
Содержание GasAlertMicro 5