GasAlertClip Extreme English Instruction Sheet
D2139-3 Page 4
Trimmed and folded dimensions: 4.125 (w) x 5.875 (h)
Calibrating the Oxygen Detector
The GasAlertClip Extreme O
2
self-calibrates using ambient oxygen levels.
Calibrate the detector
only
in a normal atmosphere (approximately 20.9%
O
2
).
To calibrate the GasAlertClip Extreme O
2
:
1. Press and hold
C
for approximately 3 seconds.
2. The detector emits one beep, one flash, and displays the following:
One vibration and a prolonged beep signifies a successful calibration. If
after a second calibration attempt the beep does not sound, contact your
BW Technologies distributor.
Gas Event Data Transmission
The detector stores the last 10 alarm events. The recorded data includes
the serial number, life remaining, self-tests performed, total number of
events and the duration of all events encountered, gas type and the alarm
level (in ppm or %), time elapsed since the alarm occurred (in days, hours,
and minutes), and the duration of the alarm (in minutes and seconds).
To transmit the data to a device:
1. With the receiving device on, position the detector and the device as
shown in the following illustration.
2. Press
C
on the detector.
3.
Prn
and
will then flash. Within 5 seconds press
C
to begin
transmission.
4. While the data is being transmitted,
Y
will come on and
will flash. A
countdown timer will display, as a percentage, how much data is left to
be transmitted.
A
Approximately 2” (5 cm)
Below is a print out example from the infrared handheld printer (GPR-
PRINTER).