BUSCH VacTest Скачать руководство пользователя страница 15

Communication | 6

Instruction Manual VACTEST DCC 400_EN_en

15 / 24

6.2.1 Setpoints

The gauge provides 2 independent, potential-free relay setpoints. These are available as
change-over switches at the connector according to the pin assignment described in
chapter 

I/O and Communication Port Schematic [

 9]

.

Relay R1, R2:

The relays can be independently configured for various switching modes. The Parameter
is used to query and set these switching modes.

Settings:

Settings

Meaning

T[p

1

] F[p

2

]

2 pressure thresholds T[p

1

] (true) und F[p

2

] (false) are transmitted. 

T[p

1

]<F[p

2

] Relay closes below p

1

 and opens above p

2

T[p

1

]>F[p

2

] Relay opens below p

2

 and closes above p

1

E

Relay closes in case of a device error.

!E

Relay opens in case of a device error.

U

Relay closes in case of pressure under-range.

!U

Relay opens in case of pressure under-range.

O

Relay closes in case of pressure over-range.

!O

Relay opens in case of pressure over-range.

C

Relay closes when cathode is switched on.

!C

Relay opens when cathode is switched on.

T0

Open relay by software command (off).

T1

Close relay by software command (on).

NOTICE

Setting T[p

1

] = F[p

2

] is forbidden!

Risk of damage to the relay!

• A gap between the thresholds being too small will result in flickering of the relay

6.2.2 Ionization Sensor

Cathode Control (CC):

It is possible to disable the ionization sensor by software command "Cathode Con-
trol" (CC) via RS485 interface:

Settings

Meaning

0

disabled 

 ionization sensor remains switched off.

1 (Default)

enabled 

 automatic ionization sensor control.

NOTE

Parameter "Cathode Control" is only temporarily saved in the gauge memory. After
mains supply is switched off or disconnected the parameter will be reset to "1" and the
ionization sensor enabled!

Cathode Control Mode (CM):

The cold cathode sensor can be switched on either manually or automatically.
The corresponding device parameter “Cathode Control Mode” (CM) can be set by soft-
ware command via RS485 interface:

Содержание VacTest

Страница 1: ...technical data instruction manuals service kits Instruction Manual VACTEST Vacuum Measurement Equipment Digital Transmitter DCC 400 0870212178 _en Original instructions Modifications reserved 16 09 2...

Страница 2: ...4 3 2 Connecting to the USB RS485 Converter 8 4 3 3 I O and Communication Port Schematic 9 4 4 Change Display Unit and Orientation Display version only 10 5 Operation 10 5 1 Before Operation 10 5 2 O...

Страница 3: ...uct Before you start working find out whether any of the vacuum components are con taminated Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts Comm...

Страница 4: ...on the product s nameplate Technical modifications are reserved without prior notification 2 3 Delivery Content Included in the delivery consignment are Digital transmitter DCC 400 Protective cover In...

Страница 5: ...pheres the device must be airproof shrink wrapped in a plastic bag together with a bag a desiccant 4 Installation WARNING Strong magnetic field Risk of severe injury The gauge contains strong permanen...

Страница 6: ...ndling the instrument make sure that the flange is protected against dirt and damage Remove the protective cover is required again during maintenance work Connect the flange to the system Connection s...

Страница 7: ...onnection cable must be used see accessories SET VacTest www buschvacuum com Active sensor controller CTR 002 CTR 004 D SUB Cables Transmitters With the controller CTR 002 up to 2 gauges can be connec...

Страница 8: ...verter Transmitter USB RS485 Converter 24 VDC Set the relevant address at the address selector switch see Setting RS485 Address 12 Connect the gauge to the power supply cable Connect the USB RS485 con...

Страница 9: ...Description Pin no Description 1 Identification 2 Signal Output 0 10 VDC 3 AGND 4 Voltage Supply 24 VDC 5 Supply GND 6 Relay 1 NO normally open 7 Relay 1 Common 8 Relay 1 NC normally closed 9 N A 10...

Страница 10: ...e Furthermore dust oil or condensing vapours will affect sensor performance and may cause malfunction NOTICE Permanent operation in the pressure range 5 x 10 4 to 2 x 10 3 mbar Wear of the cold cathod...

Страница 11: ...green LED flashing slowly Error red LED continuously on The setpoint LEDs are on when the related relay is activated For gauges with LCD display Operation error or malfunction the display is illuminat...

Страница 12: ...f ambient air 6 Communication NOTE The Busch communication protocol is available separately on request Ask your Busch representative to get the document 6 1 Setting RS485 Address The transmitter is eq...

Страница 13: ...ment range of the gauge x x x x x MV Displays the current pressure measurement x x x x x M1 Displays the current pressure of the Pirani sensor x x x x M2 Displays the current pressure of the Piezo sen...

Страница 14: ...sor Transition for more information x x x CC Configure the control of the ionization sensor See chapter Ionization Sensor 15 for more information x x x FC Configure the use of the hot cathode filament...

Страница 15: ...re over range O Relay opens in case of pressure over range C Relay closes when cathode is switched on C Relay opens when cathode is switched on T0 Open relay by software command off T1 Close relay by...

Страница 16: ...be active as long as the voltage signal is applied The parameter Cathode Control Logic DL can also be queried 6 3 VacTest Explorer Software VacTest explorer software has been especially developed for...

Страница 17: ...tion of the unit which is caused by contamination or wear and tear is not cov ered by warranty 7 1 Replacing the Sensor Head Disassembly Vent the vacuum system to atmospheric pressure and allow to coo...

Страница 18: ...8 Spare sensor Intended for DPC 400 DN 40 ISO KF flange DCC 400 DN 40 ISO KF flange DPC 400 D DN 40 ISO KF flange DCC 400 D DN 40 ISO KF flange 0680 204 599 Controller Description Part no Active Senso...

Страница 19: ...bar l s 5 x 10 10 Reaction time ms 50 Serial Interface RS485 Electrical Connection D Sub 15 pin male Supply voltage V 20 30 Cathode voltage V 2 5 Max power consumption W normal operation 3 W relays di...

Страница 20: ...Instruction Manual VACTEST DCC 400_EN_en 139 66 45 ISO KF 40 139 ISO KF 25 10 1 Gas Correction Factor Value range 0 20 8 0 Correction factor for cold cathode sensor Ar 0 80 H2 2 4 N2 1 0 Kr 0 6 CO2 0...

Страница 21: ...European Directives Electromagnetic Compatibility EMS 2014 30 EU RoHS 2011 65 EU restriction of the use of certain hazardous substances in electrical and electronic equip ment and following the stand...

Страница 22: ...Note...

Страница 23: ...Note...

Страница 24: ...nfo busch it Japan info busch co jp Korea busch busch co kr Malaysia busch busch com my Mexico info busch com mx Netherlands info busch nl New Zealand sales busch co nz Norway post busch no Peru info...

Отзывы: