10 FAULT FINDING / ERROR CODES
To cancel any error code press [ENTER] or [Esc]:
Code
Explanation
E01
Calibration has been lost.
E02
Cal difference error is too large/small.
E03
A base value with too great a difference to that currently stored.
E04
Cal was attempted with too large a value.
E05
Base was attempted with too large a value.
E06
Number of readings is greater than 999.
E07
Too many readings stored.
E08
The printer is off-line. (There may be a delay of approximately 30 seconds before receiving
this message).
E10
The memory is full.
E11
No readings were taken before trying to enter a base or cal factor.
355
Coating Thickness Gauge
29 of 48