38
von 61
38
von 61
When saving the measured values using the "Data logger" function, the comparator results are also
saved along with each measured value.
Diagram 13: Comparator
There are two different comparator modes. In the "Comparator" menu, use the
or
key to select
the comparator mode you require. Once you have made your selection, press
[ESC]
to close the
"Comparator" menu.
7.3.1. Statistic comparator mode
In static comparator mode, the comparator is stopped when the measured value goes outside the limit
window (above or below limits). The display continues to show the latest evaluation result while also
showing a box containing three exclamation marks !!!. In this case you can restart the comparator by
pressing the
[F3]
key.
Diagram 14: Measured value above comparator upper limit
7.3.2. Dynamic comparator mode
In dynamic comparator mode, the comparator evaluation continues when the measured value goes
outside the limit window (above or below limits). The current evaluation result is always displayed
without stopping the comparator.