3.11 Autoconfiguration
Autoconfiguration allows two aspects of the ForceMaster 9110 to be dealt with quickly and easily.
The ForceMaster 9110 detects the sensors that have been connected, and a teaching process
takes place in order to find the optimum monitoring values for the parts to be checked.
•
Always start setting up the ForceMaster 9110 with an autoconfiguration.
If you want to correct the monitoring values manually at a later stage, you must first carry out an
autoconfiguration.
When you start the autoconfiguration, the ForceMaster 9110 will offer you the most reasonable
monitoring mode, according to whatever sensors you have connected.
The autoconfiguration brings you to an automated result in three steps. These three steps are:
Step 1: Teaching
The connected sensors and their attachment are detected automatically, and the teaching
process is initiated. Process a workpiece. Based on this process, the ForceMaster 9110 analyses
the force-displacement profile and offers a setting for the monitoring modes and monitoring
parameters.
Step 2: Setting the modes
You can check which monitoring modes the ForceMaster 9110 suggests, and are able to specify
a different one.
Step 3: Setting the parameters
You can check what monitoring parameters have been found by the ForceMaster 9110 for the
individual monitoring regions, and have the option of correcting them.
You then change to the measuring mode. You are also able to return to Step 1, and to run
through the autoconfiguration again with other parts.
3.11.1 "Autoconfig" menu
This menu offers you a way to get into the autoconfiguration.
Autoconfig
Step 1 teach-in
Choose this menu item to start a teaching
process.
This is the usual starting point for a part that is to
be trained for the first time.
Step 2 Modes
Choose this menu item to correct the modes.
Step 3 Values
Choose this menu item to correct the values.
For a new part, start with "teach-in", and you will then be taken through all the following points.
For corrections (without a new teaching session), either choose "Step 2" to correct the modes, or
choose "Step 3" to correct the monitoring values.
If you have changed the values so much that it's no longer possible to take reasonable
measurements, then start a new autoconfiguration with "Teaching" in order to get to sensible
values again.