 Typ 9186-VX1XX

 

 

www.burster.de 

 
 

 
 

All information contained in this manual is valid for 

Typ 9186-VX1XX

 models. 

 

Typ  9186-VX1XX

 models are digital indicators fully configurables that allow input type selection in order to be 

used as needed. Available signal inputs are the following: 
 

LOAD CELL 

(30mV, 300mV) 

 

The basic instrument  consists of a soldered  assembly composed  of a main board, a display and an input  signal 
circuits. It can also be incorporated, as an option, an extra plug-in 2 SPDT 8A relays circuit output which is isolated 
from signal input and power supply. This extra circuit has independent connectors that are located on the rear part 

of the instrument once it is installed. 
 

Specially designed for direct connection to Load Cells, both models have 4 digits, configurable decimal point and 2 

LED’s for setpoints status indication, they are easy to scale into desired engineering units, directly by frontal keys 

or rear input signal value in teach mode. They also provide a 5V or 10V DC output for sensor excitation. 

 

Typ  9186-VX1XX 

models are provided with 

20mm-high 

digits and a maximum range of -

1999

 to 

9999

 which 

allows a better reading at longer distance. Minus LED sign is integrated in the most significative digit. 
 

They have three frontal keys to interact with internal software and set configuration in order to adapt their function 
to particular applications. Device programming runs through some independent menus that show short messages 
to easily identify input type and/or display configuration steps. 

 
If  relays  output  option  card  is  installed,  once  it  is  recognised  by  the  instrument,  activates  its  own  configuration 

menu which is only visible under this conditions. 

Device description 

Dimensions and mounting 

To  install  the  instrument,  prepare  a 
92x45mm  panel  cut-out  and  slide  the  unit 

inwards making sure of placing the sealing 
gasket  between  the  front  side  panel  and 

the frontal bezel. 
 
While  holding  the  unit  in  place,  put  the 

fixing  clips  on  both  sides  of  the  case  and 
slide  them  through  the  guide  tracks  until 

they reach the panel at the rear side. 

Press slightly to fasten the clips to the latching slots 
on  the  case  and  get  the  unit  fully  assembled  and 

close fitted to achieve a good sealing. 

To remove the instrument from the panel, pull outwards the rear fixing 
clips latching tabs until they are disengaged, then slide fixing clips back 

over the case. 

Содержание 9186-V Series

Страница 1: ...Typ 9186 VX1XX www burster de USER MANUAL DIGITAL INDICATOR FOR LOAD CELL INPUT Typ 9186 VX1XX...

Страница 2: ...figuration 10 DISPLAY CONFIGURATION Display programming 12 Display scaling 12 mV input 13 SETPOINTS CONFIGURATION Setpoints configuration 14 AVAILABLE KEYBOARD FUNCTIONS TARE MAX MIN and RESET functio...

Страница 3: ...ER General safety considerations Symbols identification WARNING Risk of electric shock Instrument protected by double isolation or reinforced isolation All instructions and guidelines for the installa...

Страница 4: ...brations regularly in accordance to operation criteria in each application Instrument calibration and or adjustment should be performed only by an accredited laboratory or directly by the manufacturer...

Страница 5: ...N 2 kV L N to Earth 1 kV Signal lines to Earth EN 61000 4 6 RF conducted interference Criterion A 3 Vrms EN 61000 4 11 Voltage dips 0 V during 1 cycle Criterion B 40 V during 10 12 cycles Criterion C...

Страница 6: ...mum range of 1999 to 9999 which allows a better reading at longer distance Minus LED sign is integrated in the most significative digit They have three frontal keys to interact with internal software...

Страница 7: ...nu is entered to programm the indicator whereas RUN is the normal mode in which display shows the reading according to configuration and input signal value The table below summarizes display parts des...

Страница 8: ...ls Once all connections are done plug connectors to the instrument Rear connectors location Key tool for cable insertion Load cell input signal wiring diagrams mV CN2 1 IN COMMON 2 IN 300mV 30mV 3 N C...

Страница 9: ...ween pin 4 and pin 5 of CN2 input connector must be able to provide a current of 4mA and withstand up to 40V In both cases remote TARE is activated when contact is closed less than 3 seconds Remote TA...

Страница 10: ...tion menu The instrument provides 3 keys for progressing through the menus and submenus and for data introducing modifying ENTER Vertical displacement Validates data UP Increases active digit value SH...

Страница 11: ...burster de 11 EXCITATION LEVEL Configurable value E 5 5V DC E 10 10V DC Once input range is selected the routine goes to configure excitation 5V and 10V are available In both cases load current limit...

Страница 12: ...ff frequency Fc which allows the instrument to smooth out undesirable display reading fluctuations Display scaling is necessary when adapting display reading to a particular engineering unit Display r...

Страница 13: ...isplay value indication 0000 Value entering in counts within available model display range FiLt 0 SCAL InP1 tEAC 0000 dSP1 0000 00 00 InP2 0000 dSP2 0000 For both display scaling SCAL and tEAC methods...

Страница 14: ...0 to 99 9s HyS Hysteresis in counts within available model display range 00 00 SETPOINTS CONFIGURATION If 2RE output option card is uninstalled the instrument keeps setpoints last configuration in me...

Страница 15: ...limit value defined by the working range Otherwise overflow indication OuE will be displayed Stored tare value remains in memory although power supply is removed This device detects and stores in memo...

Страница 16: ...ult this code is 0000 Finally press ENTER to begin with lock out level configuration If entered security code is wrong the instrument will go back to RUN mode 3s To access this menu from RUN mode pres...

Страница 17: ...ers a selective or total configuration lock out is available By default the unit is delivered unlocked giving access to all programming levels Once in this menu the first option will be to choose betw...

Страница 18: ...IFT key lock out for MAX MIN function is configurable in the same way as previous configurations When lock out is enabled selecting yES it is not possible to visualize maximum or minimum values by pre...

Страница 19: ...ch case just when setpoint SET is reached by display Hysteresis delay for OUT1 HI mode and for OUT2 LO mode 2RE output option allows TYP 9186 VX1XX models to perform control operations and limit value...

Страница 20: ...slide properly without much effort through rear case internal guides Each output card is supplied with an adhesive label that indicates wiring connections To help identifying terminals this label shou...

Страница 21: ...VERSION OPTION 9186 V31XX Connectors location and description 8 pin assignment for terminal implementation TARE DMS Digital Out function 9186 V31XX with Load Cell DMS wiring diagram Signal Shield Exci...

Страница 22: ...Through frontal key ENVIRONMENTAL CONDITIONS Operating temperature 10 C 60 C Storage temperature 25 C 85 C Relative humidity non condensing 95 40 C Maximum altitude 2000m Frontal protection degree IP...

Страница 23: ...template for the annotation of configured parameters in your instrument for later consulting or data recover INPUT TYPE RANGE DISPLAY CONFIG MODE INPUT 1 DISPLAY 1 INPUT 2 DISPLAY 2 FILTER 0 8 SETPOIN...

Страница 24: ...9186 VX1XX www burster de 24 burster pr zisionsmesstechnik gmbh co kg Talstr 1 5 D 76593 Gernsbach Germany Tel 49 07224 6450 Fax 49 07224 64588 Email info burster de web www burster de 28 10 2013 3072...

Отзывы: