Quick Start Guide - Dali care bed
Coupling the handset with the control unit (only in the Dali standard, wash and low-entry)
Please note: Before the bed functions can be adjusted, the handset must be coupled (connected) with the control unit.
Coupling is only necessary when commissioning the bed, replacing the Bluetooth handset and replacing the control unit.
Proceed as follows:
1. Plug the power pack into the electric socket.
• If the power pack was already in the socket, remove the power pack from the socket, wait approx. 10 s and plug it in again.
2. Within 20 s, press the two buttons for up/down mattress base height simultaneously (within 0.5 s) for at least 5 seconds.
• The LED on the control unit lights up orange for 4 seconds if coupling was successful.
• The LED on the control unit flashes orange 4 times if the coupling was not successful. In this case, repeat steps 1 and 2.
Battery
– Bluetooth handset
The Bluetooth handset is powered by a CR 2032 lithium battery.
When the battery is low, the LED display on the handset flashes yellow 4 times each time the
button is pressed. The battery must then be replaced within a few days.
Proceed as follows:
1. Open the battery compartment cover a on the back of the handset.
• For example, insert a coin into the slot of the cover and turn it counter-clockwise.
2. Remove the old battery.
3. Insert the new battery (observe polarity).
4. Close the battery compartment with the cover (turn clockwise).
Burmeier GmbH & Co. KG - Pivitsheider Straße 270 D - 32791 Lage/Lippe - Tel.: +49 52 32/98 41- 0
– Fax: +49 52 32/98 41- 41 – e-mail:
- (2019.05.24_V1_273852)
a
Turn page