Type MS08
Sensor - Parameter
38
Go to device
SAC254 Sensor
> function
Sensor
> tab
Parameter
.
Select
Sensor
.
Select the sensor type from the drop-down list: For the MS08, select
SAC
.
After changing the sensor type, the device has to be restarted.
The device restarts with the selected sensor type.
9.2
Set the interval for the automatic measurements
The parameter
Measure interval
stores the interval for the automatic measurements.
The valid range is 10 s to 86400 s (i.e. 1 day).
The default unit for the measure interval is seconds (s). However, the measure interval can also be entered
in days (d), hours (h), minutes (m), milliseconds (ms), microseconds (µs), or nanoseconds (ns).
Make sure that the login user level is
Installer
. Refer to chapter
.
Go to device
SAC254 Sensor
> function
Sensor
> tab
Parameter
.
Select
Measure interval
and enter a value.
Validate with
Apply
.
The value is set.
9.3
Adjust the measure values (customer calibration)
The parameters at
Measure values
make it possible to adjust the manufacturer calibration and configure a
customer calibration.
The customer calibration supplements the manufacturer calibration. The manufacturer calibration values are
not changed by the customer calibration. Customer calibration can be used as a fine adjustment of the
sensor for special media and is not intended to replace the manufacturer calibration.
To configure the customer calibration, 2 constants will be needed:
▪
the scaling factor or slope (default: 1)
▪
the offset (default: 0)
The calculation of the offset and the slope is explained below.
To set a customer calibration, do the following procedure:
Make sure that the login user level is
Installer
. Refer to chapter
.
Go to device
SAC254 Sensor
> function
Sensor
> tab
Parameter
.
Select
Measure values
.
Select the value you want to adjust. See Table
for the list of values available.
To set the offset, select
Offset
and enter a value. Validate with
Apply
.
To set the slope, select
Slope
and enter a value. Validate with
Apply
.
To name the adjusted value, select
Name
and enter a name. Validate with
Apply
.
Repeat for any value you want to adjust.
Value
Description
Unit of offset
SAC
Spectral absorption (254 nm)
▪
m-1 (Units per meter)
Turb 530
Turbidity (530 nm)
▪
FAU (Formazine Attenuation Units)