16
Adjustment
9.6
main menu of the configuration level
→
Press and hold the ENTER key (3 s) to access the configuration level.
The following settings are possible:
Unit
= set the unit for the flow rate, see chapter 9.6.1
InP
/
Input
= set the K factor of the fitting used, see chapter 9.6.2
Out
/
Output
= set the output signal, see chapter 9.6.3
dAtA
/
Data
= transfer of data from/to the display and configuration unit, see chapter 9.6.4
Leaving the configuration level:
End
= end the settings and go back to the Read level by pressing the ENTER key when
End
is
displayed.
9.6.1
Unit
– setting the unit for the flow rate
In this parameter, set the unit the measured flow rate is displayed in and in which the limits of the flow rate range
associated to the 4-20 mA output are defined.
When changing the unit (e.g. from liters to gallons), also change the current output limits.
Setting in the parameter:
l / s
l / m
l / h
m
3
/ m
m
3
/ h
g / s
g / m
g / s
End
Liters/second
Liters/minute
Liters/hour
Cubic meters/minute
Cubic meters/hour
US gallons/second
US gallons/minute
US gallons/hour
Status display changes to "active".
The selected unit is shown on the
display.
Fig. 11 : Unit – Setting the display for the flow rate unit