TC-Double (+) X2
Technical Data peristaltic pump CPdouble X2
Ambient temperature
0 °C to 50 °C
Flow rate
0.3 L/h (50 Hz) / 0.36 L/h (60 Hz) with standard hose
Vacuum inlet
max. 0.8 bar
Pressure inlet
max. 1 bar
Outlet pressure
1 bar
Hose
4 x 1.6 mm
Degree of protection
IP 44
Materials
Hose:
Connections:
Norprene (standard), Marprene, Fluran
PVDF
Technical Data Sample Gas Pump P1.3
Ambient temperature
0 °C to 50 °C
Operating pressure
max. 1,3 bar abs.
Nominal outlet
280 l/h (at p = 1 bar abs.)
Materials in contact with media vary
by configuration
PTFE, PVDF, 1.4571, 1.4401, Viton
AGF-PV-30-F2 Filter Technical Data
Ambient temperature
3 °C to 100 °C
max. operating pressure with filter
2 bar
Filter surface
60 cm
2
Filter mesh
2 µm
Dead volume
57 ml
Materials
Filter:
Seal:
Filter element:
PVDF, Duran glass (parts in contact with mediums)
Viton
sintered PTFE
9.3 Flow chart
Flow chart:
1 Heat exchanger 2 / pre-cooling
4 Filter (optional)
2 Heat exchanger 1 / after-cooling
5 Sample gas pump (optional)
3 Moisture detector (option)
6 Condensate pump (optional)
35
Bühler Technologies GmbH
BE440022 ◦ 10/2018