PCS.base

7.5 Drying of the moisture detector (option)

The moisture detector must be dried if moisture enters.

– Close the gas supply.

– Switch off and unplug the device.

– Loosen the swivel nut for the moisture detector connection line and disconnect the line.

– Unscrew the moisture detector counter-clockwise and remove.

– Dry moisture detector.

– Reinsert the moisture detector and carefully tighten the screw connection.

– Connect the connection line and tighten the swivel nut.

– Restore the power and gas supply.

7.6 Cleaning and removal of the heat exchanger

Heat exchangers only need to be replaced or maintained when plugged or damaged. If they are clogged, we recommend check-
ing if using a filter will avoid future occurrences.

– Close gas supply.

– Switch off device and disconnect all plugs (e.g. connector plug alarm output, supply input, etc.).

– Uninstall the device from the bag (see chapter 

Replacing the fuse of the cooler

 [> page 14]).

– Disconnect gas connections and condensate drain.

– Pull the heat exchanger up and out.

– Clean the cooling nest (hole inside the cooler block).

– Flush the heat exchanger until all contaminants have been removed.

– Grease the cooled outside surface external surface with silicone grease.

– Reinsert the heat exchanger into the cooling nest with a rotating movement.

– Reconnect the gas supply and condensate drain. The gas inlet is marked red.

– Reinstall the device in the bag.

– Restore the power and gas supply.

7.7 Replacing the hoses of the peristaltic pump (option)

– Turn off gas supply.

– Switch the device off and disconnect power supply.

– Remove the supplying and draining hoses from the pump (

Take care of the safety instructions!

).

– Loosen the centre knurled screw but do not remove it. Push the screw downwards.

– Pull off the cover.

– Pull the connections sidewards and remove the hose.

– Replace the hose and remount the pump in reverse order.

– Reconnect power supply.

7.8 Spare parts and accessories

Please also specify the model and serial number when ordering parts.

Upgrade and expansion parts can be found in our catalog.

Available spare parts:

Spare part

Item no.

Low heat device plug microfuse

5 x 20 mm, 2.5 A, delayed action

91 100 00 013

Sample gas cooler microfuse

5 x 20 mm, 5 A delayed action

91 100 00 031

Condensate vessel

CSPB 0008

15

Bühler Technologies GmbH

BE464003 ◦ 11/2017

Содержание CSPB1 series

Страница 1: ...ginal instructions Portable sample gas conditioning PCS base BE464003 11 2017 B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen Tel 49 0 21 02 49 89 0 Fax 49 0 21 02 49 89 20 E Mail analyse bueh...

Страница 2: ...this instruction carefully prior to installation and or use Pay at tention particularly to all advises and safety instructions to prevent in juries B hler Technologies can not be held responsible for...

Страница 3: ...perating the flow metre optional 10 5 4 Condensate 10 5 4 1 Version with condensate trap 10 5 4 2 Version with condensate pump 10 6 Maintenance 11 6 1 Replacing the filter element 11 7 Service and rep...

Страница 4: ...ump 115 V 2 Condensate pump 230 V When selecting the condensate trap the system can be operated at a 110 260 V voltage range 1 3 Contents PCS base in the selected version Optional accessories such as...

Страница 5: ...dify specifications without advanced notice Keep this manual for later use Signal words for warnings DANGER Signal word for an imminent danger with high risk resulting in severe injuries or death if n...

Страница 6: ...e safety and operating regulations in the respective country of use when performing any type of maintenance DANGER Electrical voltage Electrocution hazard a Disconnect the device from power supply b M...

Страница 7: ...ed in chapter Condensate page 10 DANGER Toxic corrosive condensate a Protect yourself from toxic corrosive condensate when performing any type of work b Wear appropriate protective equipment c Please...

Страница 8: ...tter of principle the operator must regard all applicable standards according prevention of damage due to lightning which may oth erwise damage the device 4 2 Connecting a gas probe The sample gas pro...

Страница 9: ...temperature is always ABOVE the output dew point setting or condensation may form in the lines after the cooler So the ambient tem perature range is limited 2 Delta T Control Here the electronics regu...

Страница 10: ...numbering of the SWs corresponds with the numbering on the DIP switch SW1 SW2 SW2 SW1 Gas output dew point 0 0 3 C 0 1 5 C factory preset 1 0 10 C 1 1 15 C SW3 SW4 SW3 SW4 Delta T control 0 0 Gas out...

Страница 11: ...ge OFF ON OK Error xxx Moisture penetration OFF Flashing f 5 Hz Error xxx xxx Various possible causes contact Service OK No error Error Error present xxx Status not defined f LED flashing frequency Al...

Страница 12: ...completely empty the ves sel as condensate could otherwise flow back into the gas lines Condensate accumulates during operation Drainage varies by version 5 4 1 Version with condensate trap Condensate...

Страница 13: ...ot be turned on unintentionally d Protect yourself during maintenance against toxic corrosive gases Use suitable pro tective equipment CAUTION Tilting risk Damage of the device Secure the device again...

Страница 14: ...point not yet reached Wait max 15 min Cooling output too long despite the cooler running Be sure the vents are not covered heat buildup Flow rate dew point gas temperature too high Maintain limits in...

Страница 15: ...health a If necessary ensure a safe gas condensate discharge b Always disconnect the gas supply when performing maintenance or repairs c Protect yourself from toxic corrosive gasses condensate when pe...

Страница 16: ...port bag Open the rear and left bag compartment Remove the two screws per see image Pull the unit out of the transport bag The fuse is located under a plastic cap on the top board see image Replace mi...

Страница 17: ...e cooler block Flush the heat exchanger until all contaminants have been removed Grease the cooled outside surface external surface with silicone grease Reinsert the heat exchanger into the cooling ne...

Страница 18: ...PVC hose DN 4 6 for gas outlet 90 14 136 Viton hose DN 4 6 for gas outlet 90 14 036 PVC hose DN 6 8 for gas inlet 90 14 138 Viton hose DN 6 8 for gas inlet 44 92 00 35 012 Condensate pump replacement...

Страница 19: ...l Dispose of parts so as not to endanger the health or environment Follow the laws in the country of use for disposing of elec tronic components and devices during disposal 17 B hler Technologies GmbH...

Страница 20: ...with condensate pump Electrical connection IEC connector Mechanical specifications Hose connections Inlet DN 6 PVDF hose nipple Output DN 4 PVDF hose nipple Weight excl accessories 6 8 kg Weight incl...

Страница 21: ...PCS base 10 Attached documents Declaration of Conformity KX460025 RMA Decontamination Statement 19 B hler Technologies GmbH BE464003 11 2017...

Страница 22: ......

Страница 23: ...ften schreiben vor dass Sie uns diese Dekontaminierungserkl rung ausgef llt und unterschrieben zur cksenden m ssen Bitte f llen Sie auch diese im Sinne der Gesundheit unserer Mitarbeiter vollst ndig a...

Страница 24: ...ce as specified above has been properly cleaned and decontaminated and that there are no risks present when dealing with the device Ansonsten ist die m gliche Gef hrdung genauer zu beschreiben In othe...

Отзывы: