GAS 222.15

2 Safety instructions

2.1 Important advice

Operation of the device is only valid if:

– the product is used under the conditions described in the installation- and operation instruction, the intended application

according to the type plate and the intended use. In case of unauthorized modifications done by the user Bühler Technolo-
gies GmbH can not be held responsible for any damage,

– when complying with the specifications and markings on the nameplates.

– the performance limits given in the datasheets and in the installation- and operation instruction are obeyed,

– monitoring devices and safety devices are installed properly,

– service and repair is carried out by Bühler Technologies GmbH,

– only original spare parts are used.

This manual is part of the equipment. The manufacturer keeps the right to modify specifications without advanced notice. Keep
this manual for later use.

Signal words for warnings

DANGER

Signal word for an imminent danger with high risk, resulting in severe injuries or death if not avoided.

WARNING

Signal word for a hazardous situation with medium risk, possibly resulting in severe injuries or death if not
avoided.

CAUTION

Signal word for a hazardous situation with low risk, resulting in damaged to the device or the property or
minor or medium injuries if not avoided.

NOTICE

Signal word for important information to the product.

Warning signs

In this manual, the following warning signs are used:

Warning against hazardous situations

General notice

Warning against electrical voltage

Disconnect from mains

Warning against respiration of toxic gases

Wear respirator

Warning against acid and corrosive substances

Wear eye/face protection

Warning against potentially explosive atmospheres

Wear protection gloves

Warning against hot surface

4

Bühler Technologies GmbH

BE460038 ◦ 10/2019

Содержание GAS 222.15

Страница 1: ...s Original instructions Sample gas probes GAS 222 15 BE460038 10 2019 B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen Tel 49 0 21 02 49 89 0 Fax 49 0 21 02 49 89 20 E Mail analyse buehler tech...

Страница 2: ...this instruction carefully prior to installation and or use Pay at tention particularly to all advises and safety instructions to prevent in juries B hler Technologies can not be held responsible for...

Страница 3: ...ting the Gas Line 8 4 5 1 Connecting the Gas Line 8 4 5 2 Connecting the calibrating gas line optional 8 4 6 Electrical connections 9 4 6 1 Plug Connection 9 5 Operation and Control 10 6 Service 11 6...

Страница 4: ...ice when connecting and the correct versions when ordering spare parts 1 2 Type Plate Example GAS 222 15 100107323 462215009902000112213 001 230V 50 60Hz 440W Material 1 4571 1 4408 PTFE Read manual M...

Страница 5: ...escription Probe Description GAS 222 15 Probe with outlet filter self regulating Accessories Please refer to the data sheet at the end of this manual for accessories for this probe 3 B hler Technologi...

Страница 6: ...odify specifications without advanced notice Keep this manual for later use Signal words for warnings DANGER Signal word for an imminent danger with high risk resulting in severe injuries or death if...

Страница 7: ...part If necessary visually inspect prior to installation to determine any obvious damage to the spare parts Always observe the applicable safety and operating regulations in the respective country of...

Страница 8: ...he original packaging or a suitable alternative The equipment must be protected from moisture and heat when not in use They must be stored in a covered dry and dust free room at a temperature between...

Страница 9: ...ary with the fitting extension must be screwed in The probe is then attached to the mating flange using the included seal and nuts 4 3 Installing the downstream filter optional NOTICE The downstream f...

Страница 10: ...line NPT 1 4 on heated probes to prevent thermal bridges Choose the shortest possible screw connection Shorten the connection pipe for the sample gas line as much as possible To do so remove the insu...

Страница 11: ...check the electric strength again yourself only do so on the respective individual components Now perform the electric strength test against earth 4 6 1 Plug Connection The probe includes two cubic p...

Страница 12: ...GAS 222 15 5 Operation and Control NOTICE The device must not be operated beyond its specifications 10 B hler Technologies GmbH BE460038 10 2019...

Страница 13: ...ic corrosive gas during maintenance Wear appropriate personal protection equipment CAUTION Hot surface Risk of burns Depending on the operating parameters the housing temperature may reach over 100 C...

Страница 14: ...he Outlet Filter with Micro Fibreglass Filter Element Turn the handle at the rear end of the probe by 90 handle must be horizontal pushing in slightly and remove Unscrew the dirty filter element count...

Страница 15: ...f the device and disconnect it from the mains b Repair the fault immediately The device should not be turned on again before elim ination of the failure Problem Malfunction Possible cause Action No or...

Страница 16: ...al Dispose of parts so as not to endanger the health or environment Follow the laws in the country of use for disposing of elec tronic components and devices during disposal 14 B hler Technologies Gmb...

Страница 17: ...ating heater 180 C Low temperature alarm Contact open at operating temperature closes at 140 C Max switching current 4 A Electrical data 230 V 2 0 A 50 60 Hz 115 V 3 8 A 50 60 Hz Max operating pressur...

Страница 18: ...GAS 222 15 9 3 Dimensions 16 B hler Technologies GmbH BE460038 10 2019...

Страница 19: ...t Mains 115 230 VAC 400 W Heater Low temperature alarm max Voltage 230 V max switching current 4 A contact opens when the operating temperature is reached contact closes when the operating temperature...

Страница 20: ...GAS 222 15 9 5 User book Please make copies Maintained on Unit no Operating hours Remarks Signature 18 B hler Technologies GmbH BE460038 10 2019...

Страница 21: ...GAS 222 15 10 Attached Documents Declaration of Conformity KX460012 Accessories Data Sheet 461099 RMA Decontamination Statement 19 B hler Technologies GmbH BE460038 10 2019...

Страница 22: ......

Страница 23: ...rs For general information see data sheet Sample gas probes GAS 222 DE461000 DE 461099 05 2021 Page 1 9 Downstream filters Cal gas connections Adapter flanges Page 5 7 Page 8 Page 2 4 B hler Technolog...

Страница 24: ...15 ANSI CSA 222 17 ANSI CSA 222 20 ANSI CSA 222 21 ANSI CSA 222 31 ANSI CSA 222 35 ANSI CSA 222 20 DH ANSI CSA 222 20 AMEX 222 21 AMEX 222 31 AMEX 222 35 AMEX 462220010300 X X X X X X X X X X X X X X...

Страница 25: ...e max 600 C 35 5 m 229 mm 600 C stainless steel 35F 0 5 m 229 mm 600 C stainless steel 07F 07 ANSI Ceramics 1 4571 Ceramics 1 4571 1000 C 1000 C 478 mm 478 mm 0 3 m 2 m 1 1 222 10 222 11 222 30 222 35...

Страница 26: ...X X X X X 462223044 X X X X X X X X X X X X X X 462223036 462223033 462223036C1 462223033C1 46222292 222 10 222 11 222 30 222 35 U 222 15 222 17 222 20 222 21 222 31 222 35 222 20 DH 222 20 Atex 222...

Страница 27: ...3 4 36 041 538 60 G3 4 36 35 229 29 G1 2 27 Beheizt heated yp L A B GF 500 40 DN65 PN6 M12 GF 1000 40 DN65 PN6 M12 GF ANSI CSA 500 40 DN3 150 M16 GF ANSI CSA 1000 40 DN3 150 M16 Unbeheizt unheated yp...

Страница 28: ...C 10 55 C 10 55 C 10 55 C 10 60 C 10 60 C 10 60 C ATEX 24 V UC II 2G D EEx m II T4 ATEX 110 V UC II 2G D EEx m II T4 ATEX 230 V UC II 2G D EEx m II T4 Bourdon tube pressure gauge 0 10 bar limit switc...

Страница 29: ...ntroller will use this input for the process logic DE 461099 05 2021 Page 7 9 A Blowback Please note that installed accessories may change the approved category of the probe Follow strictly the advice...

Страница 30: ...DE 461099 05 2021 Page 8 9...

Страница 31: ...222 30 ANSI CSA 222 35 U ANSI CSA 222 15 ANSI CSA 222 17 ANSI CSA 222 20 ANSI CSA 222 21 ANSI CSA 222 31 ANSI CSA 222 35 ANSI CSA 222 20 DH ANSI CSA 222 20 AMEX 222 21 AMEX 222 31 AMEX 222 35 AMEX 46...

Страница 32: ...aminated Ja kontaminiert mit Yes contaminated with explosiv explosive entz ndlich flammable brandf rdernd oxidizing komprimierte Gase compressed gases tzend caustic giftig Lebensgefahr poisonous risk...

Страница 33: ...rsetzen Sie vor der Inbetrieb nahme die Verkabelung wieder in den urspr nglichen Zustand Fragen Sie im Zweifel beim Hersteller nach weiteren Informationen Analysing defective assemblies is an essentia...

Отзывы: