Bühler technologies CPdouble X1 Скачать руководство пользователя страница 1

Analysentechnik

Installation and Operation Instructions

Original instructions

Peristaltic pumps

CPsingle X1, CPdouble X1

BE450026
09/2018

Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen

Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20

E-Mail: [email protected]

Internet: www.buehler-technologies.com

Содержание CPdouble X1

Страница 1: ...ginal instructions Peristaltic pumps CPsingle X1 CPdouble X1 BE450026 09 2018 B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen Tel 49 0 21 02 49 89 0 Fax 49 0 21 02 49 89 20 E Mail analyse bueh...

Страница 2: ...this instruction carefully prior to installation and or use Pay at tention particularly to all advises and safety instructions to prevent in juries B hler Technologies can not be held responsible for...

Страница 3: ...Electrical connections 6 5 Operation and control 8 6 Maintenance 9 6 1 Maintenance Schedule 9 6 2 Replacing the hose 10 6 3 Changing the pump head rotor 11 7 Service and repair 12 7 1 Safety instructi...

Страница 4: ...Peristaltic pump Product Documentation 1 3 Type plate Example CPsingle CPdouble 101001080015 44921123101 Manufacturer and address Model designation Purchase order no Item no Counter Blast protection m...

Страница 5: ...maintenance repairs and disposal must be observed These operating instructions are a part of the equipment The manufacturer reserves the right to change performance specific ation or technical data wi...

Страница 6: ...ance Repair Please note during maintenance and repairs Repairs to the unit must be performed by B hler authorised personnel Only perform conversion maintenance or installation work described in these...

Страница 7: ...e The products should be transported only in its original packaging or a suitable replacement When not in use protect the equipment against moisture and heat Keep it in a covered dry and dust free roo...

Страница 8: ...han 2 Range A per EN 60034 Cor 2010 The external potential equalisation must be established via cable at least 4 mm2 via the earthing terminal on the terminal box or the fixed and the secured metal co...

Страница 9: ...at the site must correspond with all rating plate information Protective measures against prohibited heating Unless otherwise certified the electric motor may only be used for continuous operation not...

Страница 10: ...pecifications The pump does not have a power switch It starts running as soon as the power supply is turned on NOTICE Installing peristaltic pumps CPsingle CPdouble limits the maximum permissible oper...

Страница 11: ...wear item and must regularly be checked for leaks 6 1 Maintenance Schedule Component Interval in operating hours Work to be performed To be performed by Motor After 30 000 h Replace motor Service tech...

Страница 12: ...ety notes Loosen but do not remove centre knurled nut on the hammer head screw Flip down screw Pull cover up and off Unplug external connections and remove hose Replace hose B hler spare part and inst...

Страница 13: ...rotor 2 4 between rotor 1 3 and the second bearing seat offset by 90 Then slide the shaft all the way through the rotors and the second bearing seat Guide the set screw through the bearing seat pump...

Страница 14: ...ging We will otherwise be unable to process your repair order You will find the form in the appendix of these instructions or simply request it by e mail service buehler technologies com 7 1 Safety in...

Страница 15: ...x hood single 44 9200 208Ex Ex hood double 44 9200 12Ex Ex pump head single no rotor no hood 44 9200 22Ex Ex pump head double no rotor no hood 7 2 1 Ordering information replacement hoses for 0 3 l h...

Страница 16: ...Do not dispose of the electric motor and the pressurised encapsulation as unsorted municipal waste If necessary return directly to the manufacturer Follow the laws in the country of use for disposing...

Страница 17: ...put pressure 1 bar Degree of protection IP 40 Ambient temperature 0 60 C Cord length 3 m Materials Hose Connections Norprene standard PVDF Motor markings ATEX II 2G Ex db IIB T4 Gb IECEx Ex db IIB T4...

Страница 18: ...CPsingle X1 CPdouble X1 9 2 2 Version with 2 gas paths 16 B hler Technologies GmbH BE450026 09 2018...

Страница 19: ...ouble X1 10 Attached documents Certificate IECEx IBE 17 0019 Type Examination Certificate IBExU17ATEX1008 Declaration of Conformity KX450016 RMA Decontamination Statement 17 B hler Technologies GmbH B...

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ...Untersuchungs und Pr fergebnisse werden in dem vertraulichen Pr fbericht IB 15 3 213 festgehalten 9 Die Beachtung der wesentlichen Sicherheits und Gesundheitsanforderungen wurde in bereinstimmung mit...

Страница 24: ...Hz S1 20 C bis 60 C 226S1e SG 18 9 9 0 7 5 200 240 Die genauen technischen Daten des jeweiligen Synchron Getriebemotors sind in den Pr funterlagen sowie in den der Betriebsanleitung beigef gten Datenb...

Страница 25: ...e The examination and test results are recorded in the confidential test report IB 15 3 213 9 Compliance with the essential health and safety requirements has been assured by compliance with EN 60079...

Страница 26: ...C 1 or 24 600 V AC 3 50 60 Hz S1 20 C up to 60 C 226S1e SG 18 9 9 0 7 5 200 240 The accurate technical data of the respective Synchronous gear motor are specified in the test documents as weil as in t...

Страница 27: ......

Страница 28: ......

Страница 29: ......

Отзывы: