5
MA671635_13.1
[Original Instructions]
1/23/2012
Power Switch
The main power switch is located in the left corner on the back of the machine.
To activate power to the VibroMet 2, move the power switch to the UP position.
Note:
If the VibroMet 2 malfunctions for any reason, shut the power off by moving the power
switch to the DOWN position.
Applying Polishing Cloths
The platen surface must be clean and dry before polishing cloths can be applied, especially when using
pressure sensitive (PSA) backed cloths.
1. Remove the screws securing the PVC Polishing bowl to the VibroMet 2 (see Figure 1).
2. Remove any residue from previous polishing cloths with a suitable cleaning solvent.
3. Remove the release liner on the back of the polishing cloth.
4. Place one edge of the polishing cloth on the edge of the platen.
5. Make sure the polishing cloth is centered on the platen.
6. Working in one direction, slowly press the polishing cloth on to the platen. This will help eliminate the
formation of air bubbles.
Rubbing the polishing cloth from the center out to the edge can eliminate any remaining air bubbles.
7. Replace the PVC Polishing Bowl and secure with the screws.
Figure 1 VibroMet 2 PVC Polishing Bowl
Screws securing the
PVC Polishing Bowl
PVC Polishing Bowl