7101 Presidents Dr. Suite 110, Orlando, Fl. 32809 

Phone: 800-330-2825      www.apbuck.com      Fax: 407-851-8910 

 

TechBrief 

T200-705 

BUCK BioAire™ 

Model B520 

BUCK BioAire

 Model B520 Operation 

    
   

The 

BioAire™

 Model B520 Pump provides a 

selectable, continuous, constant sample flowrate and 
is very simple to operate.  No tools are requried.  Its 
quiet operation allows unobtrusive sampling in IAQ, 
medical, clean room, public and residential building 
applications. 
   The front key pad is organized into three functional 
key groups: 

Power

 (ON-OFF), 

Timing 

(1, 2, 5, 10 

minute samples), and 

Calibration

 (Cal mode, 

Increase flow, Decrease flow). Positive tactile 
feedback is provided by the keys. The 

BioAire

  Model 

B520 Pump internal NiCad batteries provide a full five 
hours continuous run time at 15 LPM.  The standard 
charger provides a battery recharge in 12-16 hours 
and allows the unit to operate on AC power; the 
optional FastOne

 Charger provides a battery 

recharge in one hour.   
   End of sample notice is provided by the lit  LED 
“Complete” and three short beeps.

 

BioAire™

 Model B520 Sampling Steps 

     Before sampling, ensure the battery is fully charged 
and that the “Low Battery” LED is not on. For greatest 
accuracy, the 

BioAire

 Model B520 Pump should be 

calibrated once a day. 

 

1.  Place pump in desired location. 
2.  Immediately before sampling, lift up the label from 

the inlet and outlet of the cassette and attach 
them to the cassette sides.  

3.  Push the cassette (outlet side down) firmly down 

into the sample fitting on top of the BioAire

 

Model B520 Pump until it fits snugly. 

4.  Push the “ON” button to power up the  BioAire

 

Model B520 Pump and either: (a) continue to hold 
the “ON” button for four seconds to put pump in 
“Continuous” mode, or (b) push one of the 
programmed time keys (1, 2, 5, or 10 minutes) for 
an automatic timed sample. 

5.  At the end of programmed sample period, pump 

will automatically shut off and “Complete” LED will 
be lit; under “Continuous” mode, push and hold 
“OFF” button  for 2 seconds to shut off pump. 

6. 

Pull cassette up off pump with a twisting  motion 
and replace inlet and outlet stickers.

 

7.  Package the cassette(s) and send to appropriate 
        laboratory for analysis.

 

Specifications: Model B520 

Flowrate:  

5—20 LPM constant flow 

Accuracy: 

± 5% of set point 

Run Time: 

5 LPM: 20 hrs, 10 LPM: 9 hrs, 

 

    15 LPM: 5 hrs, 20 LPM: 3.5 hrs 

Size:  

    

4” H x 6” W x 5¼” D 

          

   (10.2 cm H  x  15.2 cm W  x 13.3 cm D) 

Weight:  

22 oz. (624 g) 

Compatibility:  

 

                              

 

        

 

Prices and Specifications subject to change 

  without 

notice.

 

Calibration of the 

BioAire™

 Model 

B520 Pump is fast and easy.  The supplied calibration rotameter 
assembly fits securely on top of the bioaerosol cassette with no 
leakage, so all your sample airflow goes through the bioaerosol  
cassette.  Flow    calibration with an actual cassette in-line en-
sures an accurate and repeatable calibration, without substitute 
restrictors, bubble tubes, or other fittings. 

BioAire™

 Model B520 Calibration  

     For greatest accuracy, the 

BioAire

 Model B520 Pump should 

be calibrated once per day. Calibration is performed with an 
actual bioaerosol cassette in-line, not a substitute flow restrictor 
as in other designs. The 

BioAire

 Model B520 Pump is a 

constant flow device specially designed to accommodate the 
natural backpressure inherent in 37mm bioaerosol cassettes.  It is 
for use only with such bioaerosol sampling cassettes. 
 
1.  Connect the flowmeter to the inlet of the calibration fitting.  
2.  Remove labels or caps from biaerosol cassette (retain for 

later use),  push calibration fitting down onto the top of the 
cassette  and  firmly push both pieces into the sample fitting 
on top of 

BioAire

 Model B520 . 

3.  Push “ON”, then “CAL” to put pump into calibratiion mode; 

pump will operate at last calibrated flowrate.  Adjust flowrate 
using the 

 and 

  keys to set center of rotameter ball to the 

desired flow rate  mark on rotameter.  Fine adjustments are 
possible.  Allow pump to stabilize. 

4.  When flowrate is achieved, push “CAL” to accept setting, 

then turn  pump off. 

Power on. 
Hold ON 4 sec to 
put 

BioAire

 

into 

“Continuous” 
Mode.  Hold OFF 
for 2 seconds to 
shut pump down 
 

Push one of fixed 
sample time but-
tons after Power 
“ON

” 

 

        Calibration 
            Mode 

Flow Adjust (in                 
Cal Mode) up or 
down 

 

Status LEDs  

confirm all pump 

functions  

BioAire™

 Model B520 Pump 

is  shown  with  bioaerosol    
cassette and APB-706100 
C a l i b r a t i o n  

R o t a m e t e r                 

Assembly fitted. Adjust 
flowrate on key pad under CAL 
mode up 

 or down 

 until 

center of rotameter ball is on 
desired flow rate mark. All 
sample flow calibrations are 
p e r f o r m e d  

wi t h  

a c t u a l     

bioaerosol cassette in-line. 

Rev

. 05

/27

/201

Accepts Allergenco-D and   
      Air-O-Cell

®

 Cassettes 