3 Technical data
17
R II Operation Manual, Version E
3 .2
Technical data overview
Table 3-3: Technical data
R II
R II Heating bath
Glass assemblies
A, V, C
Dimensions ( W x H x D)
550×575×415 mm
285×240×300 mm
Weight
16–18 kg (depending on the glass
assembly)
4 kg
Connection voltage
100–240 V ±10 %
100–120 V or 220–240 V ±10 %
Mains connection
3-pole (P, N, E) via power cord
3-pole (P, N, E) via power cord
Frequency
50/60 Hz
50/60 Hz
Power consumption
60 W
1700 W
Installation category
II
II
Degree of protection
IP21
IP21
Pollution degree
2
2
Rotation speed range
20–280 rpm
–
Flask size
50–4000 mL
up to 4000 mL
Max. flask content
3 kg
Temperature control
range
20 °C–95 °C
Display
Scale 0–10
Scale 0–100 °C, heating on/off
Temperature accuracy
±5 °C
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
5 – 40 °C
up to 2000 m
maximum relative humidity 80 % for temperatures up to 31 °C, and then linearly
decreasing to 50 % at 40 °C
Bath content
4 L
Heating medium
Water, distilled water (we recommend to
add borax in a concentration of 0.5 g/l when
using deionized water), heat transfer oil
Vacuum tightness of
system
5 mbar per 3 minutes at a pressure of
< 10 mbar
–
Temperature resistance
P+G
ca. -70 °C – 60 °C
–
Temperature resistance
P+G low temperature
-80 °C – 50 °C
–
Temperature resistance
protective shield
< 160 °C
–
Содержание Rotavapor R II
Страница 1: ...Operation Manual Rotavapor R II 093102 en...
Страница 2: ......
Страница 6: ...1 About this manual 6 R II Operation Manual Version E...
Страница 20: ...3 Technical data 20 R II Operation Manual Version E...
Страница 32: ...5 Putting into operation 32 R II Operation Manual Version E...
Страница 42: ...8 Troubleshooting 42 R II Operation Manual Version E...
Страница 54: ...10 Spare parts 54 R II Operation Manual Version E...
Страница 56: ...11 Declarations and requirements 56 R II Operation Manual Version E 11 2 Declaration of conformity...
Страница 57: ......