6 | Operation
BÜCHI Labortechnik AG
54/94
Operation Manual ProxiMate™
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The setting is saved.
ð
The value for the Initial Wavelength Range is saved.
6.8.7
Changing the Postpredicted Value (Calibration model only)
The following settings are available:
Postprediction type
Explanation
[Predicted Value]
Calculates the values as predicted by the chemometric
model.
[Residuum]
Is an indication of the applicability of the model.
The Residuum is the RMS value of the difference between
pre-treated sample and model reconstructed spectra.
[Mahalanobis Distance]
Calculates the value on basis of a mahalanobis calcula-
tion
[Calibration Base Stan-
dard]
Converts the parameter value of the calibration to the
value obtained at a different moisture content.
A property moisture is necessary for carrying out this post-
prediction type.
See:
Chapter 6.8.8 "Changing the Calibration Base (Calibration
model only)", page 55
Chapter 6.8.9 "Changing the Display Basis (Calibration
model only)", page 55
Proceedings:
Navigation path
➔
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
u
Navigate to the
[Application]
menu via the navigation path.
u
Tap the function
[Edit]
on the function bar.
ð
The display highlights the function
[Edit]
in green.
u
Tap the name of the application that you wish to edit.
ð
The display highlights the application in green.
u
Tap the name of the property that you wish to edit.
ð
The display shows the view
Property
.
u
Tap the action
[Postpredicted Value]
.
ð
The display shows a dialog box with the selectable values.
u
Select a value.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The setting is saved.
ð
The value is saved.