BÜCHI Labortechnik AG
Operation | 6
Operation Manual ProxiMate™
55/94
6.8.8
Changing the Calibration Base (Calibration model only)
The following settings are available:
Type
Explanation
[Dry Basis]
The property value is expressed neglecting the presence
of water in the sample.
The water contribution is subtracted from the calculation.
[Moisture xx %]
The property value is expressed assuming xx % water
content.
The amount of water is expressed as a percentage of the
total weight.
[As Is]
The property value is expressed including the presence of
water in the sample.
The water contribution is included in the calculation.
Proceedings:
Navigation path
➔
Precondition:
R
The instrument is in administrator mode. See Chapter 6.5 "Log in administrator
mode", page 39
R
The post prediction type
[Calibration Base Standard]
is selected.
u
Navigate to the
[Application]
menu via the navigation path.
u
Tap the function
[Edit]
on the function bar.
ð
The display highlights the function
[Edit]
in green.
u
Tap the name of the application that you wish to edit.
ð
The display highlights the application in green.
u
Tap the name of the property that you wish to edit.
ð
The display shows the view
Property
.
u
Tap the action
[Calibration Base]
.
ð
The display shows a dialog box with the selectable values.
u
Select a value.
u
Tap the button
[OK]
.
ð
The dialog box closes.
ð
The setting is saved.
ð
The value for the calibration base is saved.
6.8.9
Changing the Display Basis (Calibration model only)
The following settings are available:
Type
Explanation
[Dry Basis]
The property value is expressed neglecting the presence
of water in the sample.
The water contribution is subtracted from the calculation.