18
HRB Bench Scale Installation Instructions
Option installation
4.3 Column option
These instructions cover the installation of the HRB Column kits, AWT05-508878 (tall
column) and AWT05-508879 (short column). Installation for both column sizes is
identical.
Parts in the kit include the column, three screws and two adjustable feet.
To install the column, follow these steps:
1.
Lay the HRB and the column on their sides as shown in the photo below. Run
the interface cable from the indicator through the hole in the base of the column
and plug it into the proper scale connector.
2.
Place the base of the column under the bottom of the base and line up the
three holes in the column bracket to the holes in the base. Install all three
screws and tighten.
Note that the holes in the base do not have threads.
The screws will cut their own threads as they are installed.
When working on a 1 or 5kg scale,
be sure you do not set the scale
on the load platter or damage to
the load cell could occur.