![Brymen TBM161 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/brymen/tbm161/tbm161_user-manual_2823004014.webp)
13
ACA Current (Clamp-on)
RANGE Accuracy
1) 2)
400.0A
15Hz ~ 40Hz
2.0% + 5d
3)
40Hz ~ 200Hz
1.5% + 5d
200Hz ~ 400Hz @ <50A
4)
1.5% + 5d
400Hz ~ 1kHz @ <50A
4)
2.0% + 5d
800A
15Hz ~ 40Hz @ 400A ~ 600A
2.0% + 5d
3)
40Hz ~ 100Hz @ 400A ~ 600A
1.5% + 4d
15Hz ~ 60Hz @ 600A ~ 800A
5.0% + 30d
1)
Induced error from adjacent current-carrying conductor: < 0.01A/A
2)
True RMS model 162 Crest Factor: < 1.6 : 1 at full scale & < 3.2 : 1 at half scale
3)
4.0% + 5d for True RMS model 162
4)
Accuracy is specified at <50A in this frequency bandwidth due to limited calibrator
output capability for testing