6
Auto leads resistance calibration
When entering the
200
range
manually
by
RANGE
button for high precision low
resistance measurement, this feature will prompt you to short the inputs for temporary
test lead resistance calibration on this range.
The fastest way is to short the leads to auto-range to the
200
range, then press the
RANGE
button momentarily. The display sho
ws ”
Shrt
”.
Keep shorting the leads for
further 3 seconds until the display shows zero. The lead resistance is then temporarily
compensated.
The compensation value stays until the next meter power reset, and can be as high as
5
. If you need a compensation value that is higher than that, Relative Zero mode is
recommended.
Diode Test, Capacitance functions
Press the
SELECT
button momentarily to toggle the
functions. Last selection will be
saved as power up default for rep
e
at measurement convenience.