5) MAINTENANCE 

WARNING

 

To avoid electrical shock, disconnect test leads from live circuits before opening the case. Do not operate with 
open case. 

 

Cleaning and Storage 

Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is 
not to be used for periods of longer than 60 days, remove the battery and store it separately  

 

Trouble Shooting 

If the instrument fails to operate,  check battery, leads, etc., and replace as necessary. Double check operating 

procedure as described in this user’s manual

 

 

If the instrument voltage-resistance input has subjected to high voltage transient (mostly caused by lightning or 
switching surge to your system) by accident or abnormal conditions of operation, the series fusible resistors will be 
blown off (become high impedance) like fuses to protect the user and the instrument. Most measuring functions 
through this input will then be open circuit. The series fusible resistors and the spark gaps should then be replaced 
by qualified technician. Refer to the LIMITED WARRANTY section for obtaining warranty or repairing service. 

 

Battery replacement 

If  the  meter  starts  up  with  persistent  resetting  display  or  with  low  battery  icon   turns  on,  replace  the  battery 
ASAP. The meter uses one 3V coin battery IEC-CR2032.  
Before opening the case bottom, make sure the meter is switched off to avoid abrupt power reset to a running 
meter system. Disconnect test leads from live circuits. Loosen the screw from the case bottom. Lift the end of the 
case bottom nearest the input test leads until it unsnaps from the case top. Replace the battery. Observe battery 
polarities with positive (+) faces up (towards the case bottom). Replace the case bottom, and ensure that the snap 
on the case top (near the LCD side) is engaged. Re-fasten the screw.  

Note for 

Models 27s, 25s 

& 25a

 

battery replacement:  

*

Models 27s, 25s 

& 25a

 

use micro-controller (like a computer) to run the meter system. WHEN THE METER IS 

POWER-ON,  intermittence  battery  power  failure  (fast  intermittence  battery  contact  interval  in  the  order  of 
millisecond)  may  cause  the  meter  reset/re-startup  abnormally.  Simply  press  and  hold  the  SELECT  button  for 
approximately 6 seconds to master reset the system if such a situation occurs.

 

 

6) SPECIFICATION 
GENERAL SPECIFICATIONS 
Display & Update Rate:  

Model 22s

:  3-3/4 digits 4000 counts; Updates 3 per second nominal 

Models

 25a,

 25s & 27s

: 3-5/6 digits 6000 counts; Updates 5 per second nominal 

Operating Temperature:

 0

o

C ~ 40

o

Relative Humidity: 

  Maximum relative humidity 80% for temperature up to 31

o

C decreasing linearly to 50% 

relative humidity at 40

o

Altitude:

 Operating below 2000m 

Storage Temperature:

 -20

o

C ~ 60

o

C, < 80% R.H. (with battery removed)  

Temperature Coefficient:  

Nominal 0.15 x (specified accuracy)/ 

o

C @ (0

o

C ~ 18

o

C or 28

o

C ~ 40

o

C), or otherwise 

specified 

Sensing:

 Average sensing  

Pollution Degree:

 2 

Safety:

  Meets 

IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-2-030 Ed. 1.0, IEC/EN61010-2-033 Ed. 1.0, 

IEC/UL/EN61010-031 Ed. 1.1 and CAN/CSA-C22.2 No. 61010-1-12 Ed. 3.0 

Transient Protection:

 4kV lightning surge (1.2/50

s) 

Содержание BM22s

Страница 1: ... 1 Ed 3 0 IEC EN61010 2 030 Ed 1 0 IEC EN61010 2 033 Ed 1 0 IEC UL EN61010 031 Ed 1 1 and CAN CSA C22 2 No 61010 1 12 Ed 3 0 to CAT III 300V CAT II 600V CAT II 450V for Model 25s 25a In case of removing test lead tip insulation CAP to CAT II 600V CAT II 450V for Model 25s 25a only Measurement Category III CAT III is for measurements performed in the building installation Examples are measurements ...

Страница 2: ...25s 25a is present the meter displays the appropriate voltage value in DC or AC whichever larger in peak magnitude Overload Alert Feature When above rated 600V 450V for Model 25s 25a is present the meter displays OL with a warning beep tone Disconnect the test leads from the signal immediately to avoid hazards Note Range Lock Feature When a measurement reading is being displayed in AutoCheckTM mod...

Страница 3: ...s DCV is the default function Press SELECT button momentarily to select ACV The AC annunciator turns on Press momentarily again to activate the Line Level Hz function Note Line Level Hz input sensitivity varies automatically with ACV range selected when Line Level Hz is selected AC 6V range has the highest and AC 600V range has the lowest sensitivity Measuring the signal in ACV function WHILE sele...

Страница 4: ...ily AGAIN selects Capacitance function Frequency function Model 22s only Rotate the rotary selector to the Hz position selects Frequency function Unlike the Line Level Hz Frequency function for Model 27s only as stated this Common Hz Frequency function is set only at the highest input sensitivity mainly for measuring small electronic signals of below 20VAC rms 4 3 OTHER FEATURES Power On Off For M...

Страница 5: ...ect test leads from live circuits Loosen the screw from the case bottom Lift the end of the case bottom nearest the input test leads until it unsnaps from the case top Replace the battery Observe battery polarities with positive faces up towards the case bottom Replace the case bottom and ensure that the snap on the case top near the LCD side is engaged Re fasten the screw Note for Models 27s 25s ...

Страница 6: ...del 22s Data Hold and 30ms MAX Hold Model 25a 25s 27s AutoCheckTM Automatic V Selection and EF Detection Accessories Battery installed and User s manual Optional Accessories BH 20R protective holster and BSC 20 soft carrying pouch Electrical Specification Accuracy is given as of reading digits number of digits or otherwise specified 23o C 5o C and less than 75 R H DC Voltage Models 25a 25s 27s RAN...

Страница 7: ...nge accuracy is not specified 2 Accuracies with film capacitor or better 3 Updates 1 minute on large values 4 Specified with battery voltage above 2 8V half full battery Accuracy decreases gradually to 12 at low battery warning voltage of approx 2 4V Ohms Models 25a 25s 27s RANGE1 Accuracy 600 0 2 0 6d2 6 000k 1 2 6d2 60 00K 600 0K 1 0 4d 6 000M 2 0 4d Open Circuit Voltage 0 4VDC typical 1 AutoChe...

Страница 8: ...s warranty does not apply to accessories fuses fusible resistors spark gaps batteries or any product which in BRYMEN s opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation or handling To obtain warranty service contact your nearest BRYMEN authorized agent or send the product with proof of purchase and description of the difficulty postage and insura...

Отзывы: