component(s) must be disconnected from the circuit to obtain an
accurate measurement reading.
2. When using Capacitance function, discharge capacitor(s) before
making any measurements. Large value capacitors should be
discharged through an appropriate resistance load
DC
A Current function (Models 089 & 086 only)
Inputs are made through the test lead terminals
COM
/
+
.
Application notes:
The
DC
A
function is designed especially for HVAC/R flame sensor
applications. The 0.1
A resolution is useful for identifying the minute
current changes in flame detector applications. Flame signal current
check should indicate steady flame signal of at least 2
A for a
rectification type, or 1.5
A for an ultraviolet type (8
A for self
checking systems). If a flame signal current with inadequate strength
or fluctuation beyond 10%, check the following to avoid the risk of
unwanted flame relay dropout :
For gas or oil flames (Minipeeper):
Low supply voltage
Detector location
Defective detector wiring
Dirty viewing windows
Faulty Minipeeper
For oil flames (Photocell):
Detector location & wiring
Smoky flame or poorly adjusted air shutter
Faulty Photocell
Temperature over 165
o
F (74
o
C) at photocell
For gas flames (Flame Rod):
Ignition interference (A flame signal current difference with the
ignition both on and off greater than 0.5
A indicates the presence
of ignition interference)
Insufficient ground (must be at least 4 times the detector area)
Flame lifting off burner head (ground), or not continuously in
contact with the flame rod
16