6
CAUTION
Do not measure above rated frequency (1kHz @ < 50A; 100Hz @ < 800A). In
measuring above rated frequency, the magnetic circuits may reach to a hazardous
temperature
Resistance, and
Continuity functions
Inputs are made through the test leads terminals. Slide-switch on defaults at
. Press
SELECT button momentarily to select
Continuity function which is convenient for
checking wiring connections and operation of switches. A continuous beep tone
indicates a complete wire.
CAUTION
Using Resistance, Continuity or Diode function in a live circuit will produce false results
and may damage the instrument. In many cases the suspected component must be
disconnected from the circuit to obtain an accurate measurement reading
Diode test function
Inputs are made through the test leads terminals. Slide-switch on defaults at
. Press
SELECT button momentarily 2 times to select
Diode test function. Normal forward
voltage drop (forward biased) for a good silicon diode is between 0.400V to 0.900V. A
reading higher than that indicates a leaky diode (defective). A zero reading indicates a