60
Z31900_00_01
Operation Overview
8.2.3
Sample Adjustment and Maximum Sample Height
Before the spinner is inserted into the cassette:
1.
Adjust the spinner using the sample depth adjustment gauge.
2.
Insert the spinner at the
Operator Access Position
into the holder.
3.
Check if the sample (
/2) is under the maximum sample height.
The glass
tube must be below the yellow line, otherwise it will break
If the sample is too long,
do not use it
in the SampleXpress!
DANGER
Danger of injury due to glass tube breakage!
Broken glass tubes may cause minor to severe injuries as well as material damage.
Substances escaping the broken glass tubes may be hazardous or allergenic and
therefore cause minor or severe injuries or even death.
f
Wear protective equipment.
f
Refer to the corresponding precautions and cleaning/disinfection instructions.
f
Perform all tasks with the cassette and glass tubes carefully.
f
Never turn the cassette upside down or turn it on one side.
Figure 8.2
Operator Access Position
1. Maximum Sample Height
2. Sample
3. Spinner
Содержание SampleXpress
Страница 1: ...Version Bruker BioSpin think forward NMR Spectroscopy SampleXpress 001 Sample Changer User Manual ...
Страница 11: ...11 Z31900_00_01 Introduction 2 6 EC Declaration of Conformity ...
Страница 12: ...12 Z31900_00_01 Introduction 2 7 EG Konformitätserklärung ...
Страница 56: ...56 Z31900_00_01 Installation and Initial Commissioning ...
Страница 104: ...104 Z31900_00_01 Maintenance ...
Страница 116: ...116 Z31900_00_01 Error Codes ...
Страница 120: ...120 Z31900_00_01 Dismantling and Disposal ...
Страница 122: ...122 Z31900_00_01 Contact ...
Страница 144: ...144 Z31900_00_01 Avertissements De Sécurité ...
Страница 156: ...156 Z31900_00_01 Sicherheitshinweise ...
Страница 158: ......
Страница 160: ...160 Z31900_00_01 Warning Signs ...
Страница 164: ...164 Z31900_00_01 Figures ...
Страница 166: ...166 Z31900_00_01 Tables ...
Страница 168: ...168 Z31900_00_01 Glossary ...
Страница 172: ...172 Z31900_00_01 Index ...
Страница 173: ......