Advanced Operation
20
Z33085_2_002
Figure 5.2: Assign Signals to Type of Signal (Signal, Lock, Temperature).
In the next step you should import the shift-temperature values (.csv or .xml format) or fill in
the values manually. You can also create your own .xml (see below).
Figure 5.3: Dialog for importing temperature shift value files.
The .xml file (figure below) contains the name of the solvent (identical with the solvent name)
and two shift values, one for the field shift and one for the temperature shift. The values have
to be identical with the values defined for the particular solvent in
edlock
.
Figure 5.4: Example of a shift-temperature file in .xml format.
After the import the shift and temperature value are filled in the
edlock
table for the selected
solvent and used afterwards for the lock-in process Create a new XML file: