Maintenance
H148850_3_003
57
Procedure:
• Use a clean cloth to remove surface dust.
• Dampen a new cloth with the cleaning solution.
• Wipe the enclosure.
• Dry by gently buffing with a new cloth.
Avoid scratching the cover while drying. The autosampler and enclosure must be
thoroughly dry before you turn the power on.
10.3
Mounting the Z-Drive Assembly
Follow these instructions if you need to mount the Z-drive after cleaning or maintenance.
Reverse the steps if you need to remove the Z-drive.
10.3.1
Tools Required
You will need the following tools:
• Phillips-head screwdriver.
NOTICE
Device Damage Hazard
Plugging in or unplugging the device cable with the power on may cause serious damage to
the device.
u
Plug in the device cable before powering on the device.
u
Do not unplug the power cable with the device powered on.
• Turn off the sample changer.
• Slide the carriage to the end of the Y-arm.
Figure 10.1: Positioning the Carriage.
Содержание minispec
Страница 1: ...minispec Sample Automation XYZ Sample Changer User Manual Version 003 Innovation with Integrity AIC ...
Страница 6: ...Contents vi H148850_3_003 ...
Страница 10: ...Introduction 10 H148850_3_003 ...
Страница 16: ...Safety 16 H148850_3_003 ...
Страница 22: ...Transport Packaging and Storage 22 H148850_3_003 ...
Страница 54: ...Sample Stage 54 H148850_3_003 ...
Страница 66: ...Replacement of Parts 66 H148850_3_003 ...
Страница 72: ...Technical Data 72 H148850_3_003 ...
Страница 74: ...Contact 74 H148850_3_003 ...
Страница 78: ...List of Tables 78 H148850_3_003 ...
Страница 80: ...Index 80 H148850_3_003 ...
Страница 81: ...H148850_3_003 81 ...
Страница 82: ...Bruker Corporation info bruker com www bruker com Order No H148850 ...