3-4
Basic Operations and Maintenance
To power down the DektakXT system:
1
Close the Vison64 software as described in
2
Press the black O
FF
). This shuts off power to all system devices
but the computer, monitor, power supply adapter, and 24 VDC control circuit in the EMO Box.
3
Select
Start
>
Shut Down
from the Windows 7
Start
menu, and then click
Shut Down
in the dialog box
that appears.
4
Turn off the monitor.
LOADING AND UNLOADING THE SAMPLE
Loading the Sample
Click the
Unload Sample
button
on the toolbar above the Live Video Display to move the scan
stage forward 55mm (two inches). You can now conveniently place a sample on the sample fixture
(see
) remove it without risk of damage to the stylus or the sample.
Click the
Load Sample
button
to move the scan stage backward its home position. Use the
manual or auto stage controls to fine-position the area of interest on the sample beneath the stylus.
For instructions, see
Using the Manual Stage-Positioning Controls on page 3-6
and
Stage-Positioning Controls on page 3-7
Figure 3-6: Sample Loaded on the Theta Sample Fixture (Chuck)
WARNING
If the sample does not freely clear the stylus, click the Tower Up
to
raise the stylus so that it is safely above the sample. Failure to do so can
damage both the stylus and the sample.
Содержание DektakXT
Страница 1: ......
Страница 2: ...P N 980 365 DEKTAKXT STYLUS PROFILER USER MANUAL ...
Страница 8: ...Table of Contents v Applying Filters 5 11 Performing Analyses 5 12 3D Mapping Optiion 5 13 ...
Страница 9: ......
Страница 25: ......
Страница 67: ......
Страница 71: ...5 4 Using Vision64 Figure 5 4 Measurement Setup Vision View Figure 5 5 Automation Setup Vision View ...
Страница 72: ...Using Vision64 5 5 Figure 5 6 Data Acquisiton Vision View Figure 5 7 Data Analysis Vision View ...
Страница 81: ......