Lock Error Messages
100
Spectrospin AG
Daedalus Lock Manual
LOCK Wrongaddress
Error 32
The lock application software is a INTEL Hex file. If any
address in this file is bigger than the uController(80C535)
serves with its addressbus(0xFFFF) downloading will stop
immediately.
Your download file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK Wrong AppSW
Error 12
The lock boots from an EPROM. The bootroutine stored in
this EPROM gets an lock specific identifier. While the lock
powers up, the boot identifier is compared to the application
software identifier. The ”download all boards” function in the
BSMS service tool guarantees software consistency. To
restart downloading, take out jumper 6, reset the BSMS, clear
the lock errormessage #11 and start “download all boards“.
After the download has been done, the lock remains in
errorstate #11. Put in jumper 6 and reset the BSMS once
again.
LOCK WrongRecord Type
Error 33
The lock application software is a INTEL Hex file. If any line
record type specifier violates the format specifications
downloading will stop immediately.
Your download file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK WrongCheckSum
Error 34
The lock application software is a INTEL Hex file. Every line
in this file gets a checksum. The lock calculates for each line
its own checksum and compares them. If a checksum differs
from the calculated checksum downloading will stop
immediately.
Your download file is damaged, contact your local Bruker
service agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK WrongTransmissionCheck
Error 38
The lock application software is a INTEL Hex file. A stopline
indicates the end of file. If this end of file line violates the file
specification downloading will stop immediately.
Your download file is damaged, call your local Bruker service
agent for a correct LOCKxx.HEX file.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
Table 12. Lock Error Messages in alphabetical order
ERROR MESSAGE
ERROR DESCRIPTION
SIGNALS
Note: Signals of
the serial bus
(power, gain
phase, shift etc.)
are only active
during a process
change
Содержание DAEDALUS
Страница 1: ...BRUKER BRUKER SPECTROSPIN AG SERVICE MANUAL DAEDALUS LOCK Version BSMS 004 ...
Страница 6: ...Contents 6 Spectrospin AG Daedalus Lock Manual ...
Страница 14: ...8 Spectrospin AG Daedalus Lock Manual ...
Страница 16: ...10 Spectrospin AG Daedalus Lock Manual ...
Страница 19: ...Daedalus Lock Catchwords for Bash Daedalus Lock Manual Spectrospin AG 13 Figure 1 Lock Function Diagram ...
Страница 26: ...General Description 20 Spectrospin AG Daedalus Lock Manual ...
Страница 36: ...Operation 30 Spectrospin AG Daedalus Lock Manual ...
Страница 56: ...Lock Transmitter 50 Spectrospin AG Daedalus Lock Manual ...
Страница 64: ...Lock Receiver 58 Spectrospin AG Daedalus Lock Manual ...
Страница 68: ...19F RX Option 62 Spectrospin AG Daedalus Lock Manual ...
Страница 85: ...Installation Daedalus Lock Manual Spectrospin AG 79 Installation 11 2 ...
Страница 86: ...Z0 Compensation Option 80 Spectrospin AG Daedalus Lock Manual ...
Страница 90: ...Technical Data 84 Spectrospin AG Daedalus Lock Manual ...
Страница 92: ...Trouble Shooting 86 Spectrospin AG Daedalus Lock Manual ...
Страница 98: ...BSMS service tool 92 Spectrospin AG Daedalus Lock Manual ...
Страница 107: ...Error Descriptions Daedalus Lock Manual Spectrospin AG 101 ...
Страница 108: ...Lock Error Messages 102 Spectrospin AG Daedalus Lock Manual ...
Страница 114: ...Appendix 108 Spectrospin AG Daedalus Lock Manual ...
Страница 122: ...116 Spectrospin AG Daedalus Lock Manual ...
Страница 123: ...Daedalus Lock Manual Spectrospin AG 117 Lock Transmitter Index B ...
Страница 124: ...118 Spectrospin AG Daedalus Lock Manual ...
Страница 125: ...Daedalus Lock Manual Spectrospin AG 119 Lock Transmitter Index C ...
Страница 126: ...120 Spectrospin AG Daedalus Lock Manual ...
Страница 127: ...Daedalus Lock Manual Spectrospin AG 121 19F Transmitter Option Index A ...
Страница 128: ...122 Spectrospin AG Daedalus Lock Manual ...
Страница 129: ...Daedalus Lock Manual Spectrospin AG 123 Lock Receiver Index C ...
Страница 130: ...124 Spectrospin AG Daedalus Lock Manual ...
Страница 131: ...Daedalus Lock Manual Spectrospin AG 125 Lock Receiver Index D E ...
Страница 132: ...126 Spectrospin AG Daedalus Lock Manual ...
Страница 133: ...Daedalus Lock Manual Spectrospin AG 127 19F Receiver Option Index A ...
Страница 134: ...128 Spectrospin AG Daedalus Lock Manual ...
Страница 135: ...Daedalus Lock Manual Spectrospin AG 129 Lock Controller Index A ...
Страница 136: ...130 Spectrospin AG Daedalus Lock Manual ...
Страница 137: ...Daedalus Lock Manual Spectrospin AG 131 Lock Controller Index B C ...
Страница 138: ...132 Spectrospin AG Daedalus Lock Manual ...
Страница 139: ...Daedalus Lock Manual Spectrospin AG 133 Lock RS232 Biggy Board ...
Страница 140: ...134 Spectrospin AG Daedalus Lock Manual ...
Страница 141: ...Daedalus Lock Manual Spectrospin AG 135 Z0 Compensation Option ...
Страница 142: ...136 Spectrospin AG Daedalus Lock Manual ...
Страница 143: ...Daedalus Lock Manual Spectrospin AG 137 Cables ...
Страница 144: ...138 Spectrospin AG Daedalus Lock Manual ...
Страница 145: ...Daedalus Lock Manual Spectrospin AG 139 Test Equipment Index A ...