Error Descriptions
Daedalus Lock Manual
Spectrospin AG
95
LOCK ErrorOnLCB
Error 4
This is displayed if the CPU, Keyboard or X32 ignore an
error that is detected by the Lock and sent to the CPU.
LOCK Error_Startchar
Error 30
The lock application software is a INTEL Hex file. If the
starting character of any line in this file violates the format
specifications downloading will stop immediately.
To restart the download, reset the BSMS and clear the LOCK
Errormessage #11.(The lock uses only the booteprom, no
valid application software in its program memory).
LOCK FFAError
Error 143
Appears if the DSP FFA is not correctly completed within a
certain time period.
Possible Causes:
- Signal ADC_CONV (L-RX) is not available
- ADC (IC9) on the lock receiver is defect.
LOCK GainBusError
Error 130
Displayed if RF Gain could not be loaded. The
microprocessor on the lock controller board loads the D/A
converter in the Lock Receiver (IC10) via a serial Bus during
a Gain-setting. The microprocessor reads the digital Gain
value back in and compares it with the transmitted value. If
the values are not the same a Gain Bus Error is generated.
WR
~
CLK
CS_Gain
DATA
DATA_RETURN
LOCK HF Power Supply Failure
Error 157
All analog power supplies (LOCK_P15V, LOCK_N15V,
LOCK_P5V and LOCK_N5V) are checked by the
Microcontroller. The supplies are multiplied by a individual
factor and added to a common signal L-RX, IC18).
Possible Causes:
- One or several Supplies missing.
- Some Power Supplies do not fulfil their specifications.
MUX L-RX (IC19) Pin
10
LOCK_P15V
LOCK_N15V
LOCK_P5V
LOCK_5V
LOCK HWTesterIsOff
Error 50
Function is only available in Hardware Testmode.
LOCK HWTesterIsOn
Error 51
System is not available as the Hardware Testmode is active.
Switch to the normal Mode with ‘Reset’.
LOCK ModeBusError
Error 134
Appears if the Mode could not be loaded. The microprocessor
on the LCB loads the transmitter shift register (IC27) via a
serial Bus using a Mode change (e.g. FFA and Normal Lock
Mode). The mode settings are read into the microprocessor
again and compared with the transmitted values. If the values
differ, a Mode Bus Error is generated.
WR
~
CLK
CS_CNTR
DATA
STATUS_DATAR
LOCK No Communication with
DSP
Error 149
This error appears, if the DSP does not clear its host interface
receive registers. Normally, this registers are released after a
read cycle by the DSP semiconductor
Possible Causes:
- Protocol Failure at the Host Bus.
- Control Signals for Host Bus are not generated correctly.
LCB:
DSP_HEN~
DSP_HR_W~
uP: A0..A2
uP: D0..D7
Table 12. Lock Error Messages in alphabetical order
ERROR MESSAGE
ERROR DESCRIPTION
SIGNALS
Note: Signals of
the serial bus
(power, gain
phase, shift etc.)
are only active
during a process
change
Содержание DAEDALUS
Страница 1: ...BRUKER BRUKER SPECTROSPIN AG SERVICE MANUAL DAEDALUS LOCK Version BSMS 004 ...
Страница 6: ...Contents 6 Spectrospin AG Daedalus Lock Manual ...
Страница 14: ...8 Spectrospin AG Daedalus Lock Manual ...
Страница 16: ...10 Spectrospin AG Daedalus Lock Manual ...
Страница 19: ...Daedalus Lock Catchwords for Bash Daedalus Lock Manual Spectrospin AG 13 Figure 1 Lock Function Diagram ...
Страница 26: ...General Description 20 Spectrospin AG Daedalus Lock Manual ...
Страница 36: ...Operation 30 Spectrospin AG Daedalus Lock Manual ...
Страница 56: ...Lock Transmitter 50 Spectrospin AG Daedalus Lock Manual ...
Страница 64: ...Lock Receiver 58 Spectrospin AG Daedalus Lock Manual ...
Страница 68: ...19F RX Option 62 Spectrospin AG Daedalus Lock Manual ...
Страница 85: ...Installation Daedalus Lock Manual Spectrospin AG 79 Installation 11 2 ...
Страница 86: ...Z0 Compensation Option 80 Spectrospin AG Daedalus Lock Manual ...
Страница 90: ...Technical Data 84 Spectrospin AG Daedalus Lock Manual ...
Страница 92: ...Trouble Shooting 86 Spectrospin AG Daedalus Lock Manual ...
Страница 98: ...BSMS service tool 92 Spectrospin AG Daedalus Lock Manual ...
Страница 107: ...Error Descriptions Daedalus Lock Manual Spectrospin AG 101 ...
Страница 108: ...Lock Error Messages 102 Spectrospin AG Daedalus Lock Manual ...
Страница 114: ...Appendix 108 Spectrospin AG Daedalus Lock Manual ...
Страница 122: ...116 Spectrospin AG Daedalus Lock Manual ...
Страница 123: ...Daedalus Lock Manual Spectrospin AG 117 Lock Transmitter Index B ...
Страница 124: ...118 Spectrospin AG Daedalus Lock Manual ...
Страница 125: ...Daedalus Lock Manual Spectrospin AG 119 Lock Transmitter Index C ...
Страница 126: ...120 Spectrospin AG Daedalus Lock Manual ...
Страница 127: ...Daedalus Lock Manual Spectrospin AG 121 19F Transmitter Option Index A ...
Страница 128: ...122 Spectrospin AG Daedalus Lock Manual ...
Страница 129: ...Daedalus Lock Manual Spectrospin AG 123 Lock Receiver Index C ...
Страница 130: ...124 Spectrospin AG Daedalus Lock Manual ...
Страница 131: ...Daedalus Lock Manual Spectrospin AG 125 Lock Receiver Index D E ...
Страница 132: ...126 Spectrospin AG Daedalus Lock Manual ...
Страница 133: ...Daedalus Lock Manual Spectrospin AG 127 19F Receiver Option Index A ...
Страница 134: ...128 Spectrospin AG Daedalus Lock Manual ...
Страница 135: ...Daedalus Lock Manual Spectrospin AG 129 Lock Controller Index A ...
Страница 136: ...130 Spectrospin AG Daedalus Lock Manual ...
Страница 137: ...Daedalus Lock Manual Spectrospin AG 131 Lock Controller Index B C ...
Страница 138: ...132 Spectrospin AG Daedalus Lock Manual ...
Страница 139: ...Daedalus Lock Manual Spectrospin AG 133 Lock RS232 Biggy Board ...
Страница 140: ...134 Spectrospin AG Daedalus Lock Manual ...
Страница 141: ...Daedalus Lock Manual Spectrospin AG 135 Z0 Compensation Option ...
Страница 142: ...136 Spectrospin AG Daedalus Lock Manual ...
Страница 143: ...Daedalus Lock Manual Spectrospin AG 137 Cables ...
Страница 144: ...138 Spectrospin AG Daedalus Lock Manual ...
Страница 145: ...Daedalus Lock Manual Spectrospin AG 139 Test Equipment Index A ...