BSMS/2 Mainframe Version 001
Bruker
33
Installation
5.2
To install the keyboard please follow the guide below, making sure to do each step in the
order given. No hardware adjustments are necessary for the BSMS keyboard!
1.
Connect the BSMS keyboard to the CPU using a standard 25 pin cable (22887).
2.
Switch on the BSMS. Check that the messages
‘Booting’
and afterwards
‘Con-
necting
…
’
appear on the keyboard display. After approximately one second,
also verify that the green KEYB LED on the CPU is blinking. No error LED’s
on the CPU should be lit and no error message from the keyboard (KBC) should
appear on the display.
3.
Check the software version with the BSMS Service Tool and download the new
software if necessary.
Keyboard Controller Board
5.3
The keyboard controller board is located in the bottom of the BSMS keyboard. It is
a processor system built around a 80C535 micro-controller. The application soft-
ware runs on a real time operating system and can be downloaded via the serial link
with the CPU (see BSMS Service Tool Manual). The control knob and the keyboard
display board are connected to the keyboard controller board.
Keyboard Display Board
5.4
All keys, LED’s with drivers, and the 2
×
8 alphanumeric display of the BSMS key-
board are placed on the keyboard display board. It is connected with the keyboard
controller board by a 64 wire flat cable.
Содержание BSMS/2
Страница 1: ...BRUKER BSMS 2 BSMS 2 Mainframe Technical Manual Version 001 Bruker ...
Страница 28: ...Central Processing Board CPU 28 Bruker BSMS 2 Mainframe Version 001 ...
Страница 36: ...Buses and Wiring 36 Bruker BSMS 2 Mainframe Version 001 Chassis wiring 6 2 ...
Страница 37: ...BSMS 2 Mainframe Version 001 Bruker 37 User Bus wiring 6 3 ...
Страница 38: ...Buses and Wiring 38 Bruker BSMS 2 Mainframe Version 001 JU1 6 are placed placed ...
Страница 39: ...BSMS 2 Mainframe Version 001 Bruker 39 ...
Страница 40: ...Buses and Wiring 40 Bruker BSMS 2 Mainframe Version 001 ...
Страница 41: ...BSMS 2 Mainframe Version 001 Bruker 41 ...
Страница 42: ...Buses and Wiring 42 Bruker BSMS 2 Mainframe Version 001 ...
Страница 46: ...BVT 46 Bruker BSMS 2 Mainframe Version 001 Attach the cooling air channel and tighten the screws ...
Страница 47: ...Chapter BSMS 2 Mainframe Version 001 Bruker 47 Schemata 8 Power supply PSB1 place plan 8 1 ...
Страница 48: ...Schemata 48 Bruker BSMS 2 Mainframe Version 001 Power Supply PSB2 place plan 8 2 ...
Страница 49: ...BSMS 2 Mainframe Version 001 Bruker 49 Variant Scheme PSB1 PSB2 8 3 ...
Страница 50: ...Schemata 50 Bruker BSMS 2 Mainframe Version 001 ...
Страница 51: ...BSMS 2 Mainframe Version 001 Bruker 51 ...
Страница 52: ...Schemata 52 Bruker BSMS 2 Mainframe Version 001 ...
Страница 53: ...BSMS 2 Mainframe Version 001 Bruker 53 ...
Страница 54: ...Schemata 54 Bruker BSMS 2 Mainframe Version 001 ...
Страница 55: ...BSMS 2 Mainframe Version 001 Bruker 55 ...
Страница 56: ...Schemata 56 Bruker BSMS 2 Mainframe Version 001 ...
Страница 57: ...BSMS 2 Mainframe Version 001 Bruker 57 ...
Страница 58: ...Schemata 58 Bruker BSMS 2 Mainframe Version 001 ...
Страница 59: ...BSMS 2 Mainframe Version 001 Bruker 59 ...
Страница 60: ...Schemata 60 Bruker BSMS 2 Mainframe Version 001 ...
Страница 61: ...BSMS 2 Mainframe Version 001 Bruker 61 Lastpage ...