77
Z31823_00_05
Troubleshooting
11.1.1.33
Pump Test Pump 2 Failed
The automatic pump test detected that the pump P2 does not work reliable.
Solution:
Check the pump very careful. If it is not working correctly, exchange the pump
with the spare pump. To change the pump read the section
11.1.1.34
Pump Test Not Possible, Parameter Error
The automatic pump test could not be started because the parameters for helium flow
„Fl“ and/or pressure1 „P1“ could not be read correctly.
Solution:
Read the section
"When the Value of a Parameter is Displayed as NAN" on
.
11.2
Messages in the Event Log
Any messages that appear during operation of the BMPC/2 are displayed In the Event
Log of the „Present Values/Events“ window. In addition the messages are stored in the
Event Log file.
There are several different kinds of messages, which can be displayed or faded out
using the filter adjustment of the event log browser:
•
Warning
: The upper or lower limits of a warning limit have been surpassed.
•
Alarm
: The upper or lower limits of an alarm limit have been surpassed.
•
Misc
: A general message, e.g. change of mode.
•
Control Unit
: All messages are transferred from the BMPC/2 through the serial
port. These error message are marked with the prefix „CU“.
11.2.1
Additional Error Messages in the Event Log
CAN Network Reboot
CAN Slave General
CAN Slave No Data
CAN Slave Initialization Failed
CAN Slave Unexpected Answer
CAN Slave Hardware Error
CAN Slave SBS Error
CAN Slave Reset Event
CAN Fatal RTX Initialization Failed
CAN Fatal RTX Internal
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......