50
Z31823_00_05
Hardware Overview
8.2.4
UPS
Uninterruptable Power Supply (ECL < 01.00: P/N O10440;
.00: P/N 1802515)
•
Input: 208V, 200V*, 208V**, 220V, 230V (default), 240V
*100V and 200V are derated by 20%
**110V and 208V are derated by 10%
Voltage range at 100%: 180-276 Vac.
•
1000VA
•
Bridging time: 7 minutes at full output load (1000VA)
•
19“ rack mount 2HE
For more details refer to the separate UPS manual.
8.2.5
Pump Cabinet
The pump cabinet consists of:
•
2 vacuum pumps (P1 and P2) and 2 valves (V1 and V2), connected to the pumping
line.
•
A helium flow meter which measures the helium flow at the exhaust of both pumps.
The flow meter (P/N 87487) is connected to the BMPC/2 control unit via the CAN
bus.
•
A temperature sensor (PT100) to measure the temperature T5 inside the pump cab-
inet.
ECL
01
≥
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......