Alarm System
6.2
Alarm Configuration
All channels that have the option „Alarm Trigger“ checked can trigger an alarm cycle.
The alarm trigger option can be configured using the function from the menu
Tools -
Options - Sensor Channels
.(for additional information see
).
An alarm will also be triggered if the communication between the industrial PC and the
BMPC/2 Control Unit (RS232) fails. This may occur, for example, when a power failure
occurs, which lasts longer then the backup time of the external UPS, or which is not pro-
tected by the external UPS. The PC then will be powered for 30 more minutes using the
internal UPS, whereas an alarm cycle can be initiated during this time.
6.3
Alarm Sequence
When the system detects a deviation above or below a limit value, then an alarm cycle
will be started.
1. After 30 seconds a check will be made to see if the deviation still exists. When the
deviation still exists, the alarm cycle will continue (start calling), otherwise, the alarm
cycle will be terminated.
No Deviation
: Alarm cycle will be terminated
2. When the cycle continues the people from the
(see
) are called sequentially. To learn what you have to do when
you receive an alarm call, please see
Acknowledgement
: Alarm calls stopped. When the person who has assumed
responsibility for the system (by acknowledging the alarm call) switches the sys-
tem on site to „OFF“ mode, a confirmation fax will be sent to all alarm fax recipi-
ents to show that the magnet system is under control.
3. When no acknowledgement takes place the
Acknowledgement
: Alarm calls stopped. When the person who has assumed
responsibility for the system (by acknowledging the alarm call) switches the sys-
tem on site to „OFF“ mode, a confirmation fax will be sent to all alarm fax recipi-
ents to show that the magnet system is under control.
4. When the fax group is enabled, a fax will be sent to the entire
5. When this option is enabled, an e-mail will be sent to the entire
.
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......